Page 2
83 to 264 V ac, 45 to 65 Hz, 100 to 375Vdc, 35W/70 VA On page 6-5, under Flexible Current Probes for AC, replace the Model No. Product No. column with: Model No. Product No. TPS FLEX 18-TF-II 4531837 TPS FLEX 24-TF-II 4531843 TPS FLEX 36-TF-II...
Page 3
Manual Supplement 1760 Users Change #2, 67378, 67383 On page 1-3, in the Symbols table, replace the CAT II, III, IV entires with: Conforms to relevant South Korean EMC Standards. Ã MEASUREMENT CATEGORY II is applicable to test and measuring circuits CAT II connected directly to utilization points (socket outlets and similar points) of the low- voltage MAINS installation.
Page 4
1760 Users Manual Supplement On page 1-8, replace Figure 1-3 with: Grafikview.eps Figure 1-3. Top View On page 1-9, replace Table 1-2, with: Item Description Mains connection Mains switch LED indicators Ethernet connector USB connectors ...
Page 5
Manual Supplement 1760 Users Under 2 USB connectors, replace the Note with: Note A USB stick must not be pulled out while the USB LED flashes. This can cause permanent damage to the storage media. Only remove the USB stick while the USB LED is OFF or statically ON.
Page 6
1760 Users Manual Supplement Change #3, On page 6-4, in the first table, replace footnotes 2 and 3 with: sensor + basic unit 0.1 % of Udin = 230 V P-N in accordance with IEC61000-4-30 Class A - for values <10 % and >150 % of Udin 0.15 % of rdg ±0.48 V sensor + basic unit 0.1 % of Udin = 480 V and 600 V P-N in accordance with IEC61000-4-30 Class A - for values <10 % and >150 % of Udin 0.15 % of rdg ±0.8 V 2/18...