Page 2
Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available only if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price.
Table of Contents Chapter Title Page Function ....................1-1 General Function....................1-1 How to Contact Fluke ..................1-1 Safety Information ..................... 1-2 Symbols Used in this Manual ................1-2 Components ....................... 1-2 PC Requirements ....................1-2 LEM Basics......................1-3 Calibration Data Storage ................1-3 Main Comm Port ...................
Page 4
RUSKA 2456-LEM Users Manual Intervals ......................2-7 File Names..................... 2-7 Default Location.................... 2-7 File Verification..................... 2-7 Multiple LEMs....................2-8 Cabling ......................2-8 Addressing..................... 2-8 RUSKA 2456 Operation with the LEM..........3-1 Communications ....................3-1 Winprompt Interface..................3-1 RUSKA 2456 and 2465 Upgrade for LEM Compatibility ........ 3-1 Remote Pressure Capture ..............
Page 5
Contents (continued) Remote Commands.................... 9-1 DruckBus Binary Standard................9-1 DruckBus Compatibility Standard..............9-1 Terminology: ....................9-1 Timing ......................9-2 Global Command Responses................. 9-2 DruckBus Binary Mode Commands.............. 9-2 @ - Set Address..................... 9-2 A – Read ADC Counts .................. 9-3 C –...
List of Tables Table Title Page 1-1. Symbols........................1-2 4-1. RUSKA 6220 25 Pin RS-232 Connector ............... 4-3 4-2. RUSKA 6220 DB25 and LEM DB9 Connections ..........4-3 6-1. DB9 Pinout......................6-1...
List of Figures Figure Title Page 1-1. LEM General View ....................1-3 1-2. LEM Communication Switch................. 1-4 2-1. Serial Port Setup..................... 2-2 2-2. View Coefficients....................2-2 2-3. Options ........................2-3 2-4. Units ........................2-3 2-5. Limits ........................2-4 2-6. Select Password...................... 2-5 2-7.
RUSKA 7220 / 6220 Portable Pressure Gauges and other pressure standards. How to Contact Fluke To order accessories, receive operating assistance, or get the location of the nearest Fluke distributor or Service Center, call: Technical Support USA: 1-800-99-FLUKE (1-800-993-5853)
AC (Alternating Current) Earth Ground Important Information: refer to manual Do not dispose of this product as unsorted municipal waste. Go to Fluke’s website for recycling information. Components The LEM product comes with the following accessories: Power Plug (input: 110-240 Vac, 50/60Hz) Part #62-315...
Function LEM Basics LEM Basics Local Pressure Port Temperature Sensor Serial Pressure Input Humidity Sensor Main Comm. Port Power Indicator Power Input gky01.eps Figure 1-1. LEM General View Calibration Data Storage The calibration data is stored in the LEM. Refer to Chapter 5 for information on the Calibration process and how to access the coefficients.
RUSKA 2456-LEM Users Manual RS232 - C1 RS458 - C2 gky02.eps Figure 1-2. LEM Communication Switch Optional USB cables are available: Part # 8-825 - USB to RS232 (requires Windows 98 or higher). Power The Power LED will illuminate when power is applied through either the power plug or the main communication port.
Chapter 2 Getting Started Introduction The following explains how to setup PC operation with RUSKA LEMCal software and a single LEM. Interconnect The standard interface cable for the LEM is a DB9 F/F cable. The green indicator light should come on when the power plug is connected. Software Installation 1.
RUSKA 2456-LEM Users Manual gky04.bmp Figure 2-1. Serial Port Setup The default settings are shown. Change the Comm Port to match the host computer. The Baudrate, Databits, Parity and Stopbit defaults should be set as shown in Figure 2-1. Communications Protocol See Appendix, Remote Commands, for a description of the Communication Protocol.
Getting Started Software Setup LEMCAL Software The factory default address is 33. The LEM and LEMCAL Software default to this Address. The LEMCAL Software needs the address set in the SETUP/OPTIONS screen of the LEM with which it will communicate. If this address is set to 0 then all LEMs on the communication bus will respond.
RUSKA 2456-LEM Users Manual gky08.bmp Figure 2-5. Limits Select values for the upper and lower limits for Pressure, Temperature, and Relative Humidity. Setup The Serial Number for the desired LEM is selected in the box in the top, left corner.
Getting Started Software Setup gky09.bmp Figure 2-6. Select Password The password protection will prevent the changing of the calibration data and address. The default password is “0” (zero). When the password is “0” the password protection is turned off. When setting the password for the first time, “0” must be entered as the Old Password. The password can be reset to “0”...
RUSKA 2456-LEM Users Manual Viewing Real Time Data gky11.bmp Figure 2-8. Main Screen The Memo line allows a user to enter a descriptive to identify each LEM. This description is associated with the Serial Number and is stored in the *.ini file on the host computer.
Getting Started Logging Data gky12.bmp Figure 2-9. Data Logging The Module number, in the left hand column, is referenced to each LEM as defined in Figures 2-3 and 2-62-6. To begin logging data for a particular LEM, select the Enable button.
RUSKA 2456-LEM Users Manual Select the *.LOG file that needs to be verified. Upon completion of the verification a notice of verification will be displayed or a file *.ERR will be created and saved in the LOG directory. This file will identify which lines of the *.LOG file have been corrupted or altered.
Chapter 3 RUSKA 2456 Operation with the LEM Communications Set the communications interface to RS485 for use with the 2456 and 2465. Refer to Chapter 1, Main Comm Port, for LEM communication setup. Power for the LEM is supplied from the 2456 or 2465, over the communications cable. Use a standard DB9M / DB9F cable.
Chapter 4 Remote Pressure Capture Remote Pressure Capture From the Menu select Setup | Options (refer to Figure 4-1). Set the Pressure Source as Remote. Click the extension arrows to the right of the selected LEM. The Remote String Dialog box will appear.
RUSKA 2456-LEM Users Manual Use with a RUSKA Portable Pressure Gauge 7220 or 6220 gky14.bmp Figure 4-2. Remote String Dialog Initialization String Transmitted from LEM to remote pressure sensor to initialize the remote sensor. Pressure String Transmitted from LEM, on a continuous basis, to retrieve pressure readings.
Remote Pressure Capture Cabling Cabling Connecting the RUSKA 6220 to the LEM Pressure Serial Port Table 4-1. RUSKA 6220 25 Pin RS-232 Connector DESCRIPTION Transmitted data (to PPG) Receive Data (from PPG) RTS (must be set for PPG to transmit) CTS (always set by PPG) *Connect Pins 4 &...
RUSKA 2456-LEM Users Manual Scaling Once the raw value is retrieved from the string then units or a scaling factor must be given. Refer to Chapter 4, Remote Pressure Units. Communication Port Configuration This configuration applies only to the Remote pressure Input and is set as follows:...
Chapter 5 Calibration Span Pressure - On Board Sensor Calibration Point 1: 10.5 +/-0.25 psia Calibration Point 2: 15.0 +/-0.25 psia Temperature Calibration Point 1: 17 +/- 2 degC Calibration Point 2: 29 +/- 2 degC Zeroing Temperature measuring devices are impacted by self heating. The rise in temperature, measured by the sensor, is a function of the amount of heat generated by the probe and the ability of the probe to dissipate this heat.
RUSKA 2456-LEM Users Manual Calibration Data In order to view the calibration coefficients, select Calibrate | View Coefficients. gky15.bmp Figure 5-1. View Coefficients The menu bar selections allow the coefficients to be read and written to disk and read and...
Chapter 6 Connector Definition Power Plug The power plug is a barrel type receptacle: ID = 2.5 mm, OD = 5.5 mm, Depth = 10.0 mm, Inside - Positive DB9 Connectors Table 6-1. DB9 Pinout DB9 male DB9 male DB9 female Pin #s DB9 RS 485 RS 232...
Chapter 7 Air Density Calculation LEM Internal Calculation vs. RUSKA PC Software Calculation The Air Density calculation available directly from the LEM is a best fit curve approximation as described below in Chapter 7, Air Density Available Directly from LEM. The Air Density output available from RUSKA PC Software (WinPrompt, LEM PC, etc.) uses the definition available per National Bureau of Standards, NBS Special Publication 700-1, Industrial Measurement Series (November 1984).
Page 36
RUSKA 2456-LEM Users Manual Data Update Rate: New Data available at 1 second intervals. Includes Temperature, Humidity, Pressure, Air Density. Available Units: from Module : ºC x100, %RH x100, kPa x100, g/m from *.dll: ºC, %RH, kPa, g/m from PC Software : ºC, ºF;...
Chapter 9 Remote Commands Remote Commands There are two general standards supported. The Binary protocol requires fewer bytes, therefore requiring less time to transmit each message. The Compatibility protocol requires more bytes per message, but allows the module to co-exist on the same RS485 bus with other DruckBus compatible components. DruckBus Binary Standard Command: &<Addr><Size><Cmd>[<Parm1>[<Parm2>...]]<Chk>...
RUSKA 2456-LEM Users Manual D – Read Air Density Command: <start><addr><Size>D<chk> Reply: <start><addr><Size>d<status><density><chk> Command <Size>: 1 Reply <Size>: 6 <status> is three unsigned 8-bit values. (See Command A) Returned data is a 16-bit unsigned value, least significant byte first. E – Read EEPROM Command: <start><addr><Size>E<addr1><chk>...
Remote Commands Remote Commands R - Read ADC Reference Values Command: <start><addr><Size>R<chk> Reply: <start><addr><Size>r<temp><rh><press><chk> Command <Size>: 1 Reply <Size>: 7 All values are 16-bit signed integers using scaled (*100) units. <temp> is degrees C <rh> is % <press> is kPa. S - Read Remote Pressure with Displayed Precision Command: <start><addr><Size>S<chk>...