Thermocouple Measurement Range Accuracy Test - Fluke 2620A User Manual

Hydra series ii data acquisition unit, hydra series ii data logger
Hide thumbs Also See for 2620A:
Table of Contents

Advertisement

Table 6-3. Performance Tests (Voltage, Resistance, and Frequency) (cont)
Using inputs in decades of 1.9:
300
3 k
30 k
300 k
3 M
Using inputs in decades of 1:
300
3 k
3 k
30 k
300 k
3 M
10 M *
*Optional test point if standards available.
All channels (0 through 20) can accommodate 2-terminal resistance measurements. Channel 0, with only
two connections, cannot be used for 4-terminal measurements. Four-terminal resistance measurements
can be defined for channels 1 through 10 only. Channels 11 through 20 are used, as required, for 4-
terminal to provide the additional two connections. For example, a 4-terminal set up on channel 1 uses
channels 1 and 11, each channel providing two connections.
Frequency
90 kHz

Thermocouple Measurement Range Accuracy Test

Assure the Accuracy Verification Test for channel 0 meets minimum acceptable levels
before performing this test.
Thermocouple temperature measurements are accomplished using Hydra Series II's
internal 100 mV and 1V dc ranges. (The ranges are not configurable by the operator.)
This procedure will provide the means to test these ranges.
To test the 100 mV and 1V dc ranges requires computer interfacing with a host (terminal
or computer). The host must send commands to select these ranges. These ranges cannot
be selected from the front panel of Hydra Series II.
1. Ensure that communication parameters ( i.e., transmission mode, baud rate, parity,
and echo mode) on the Hydra Series II and the host are properly configured to send
and receive serial data. Refer to Chapter 4.
2. Power up Hydra Series II and allow it to temperature stabilize for 1/2 hour.
short
190
short
1.9 k
19 k
190 k
1.9 M
short
100
short
1 k
10 k
100 k
1 m
10 M
Note
10 kHz/2V p-p
Performance Tests
0.00
0.09
189.93
190.12
0.0000
0.0003
1.8992
1.9008
18.992
19.008
189.91
190.09
1.8983
1.9017
0.00
0.09
99. 95
100.10
0.0000
0.0003
0. 9995
1.0006
9. 995
10.005
99. 94
100.06
0. 9990
1.0010
9. 979
10.021
9.994
10.006
Maintenance
6
6-9