Table 4-8. Command and Query Reference (cont)
Service Request Enable
Sets the Service Request Enable Register to the given value.
*SRE <value>
<value> = (0 .. 255)
If the value is greater than 255, a Command Error is generated. The
value of bit 6 is ignored, since it is not used by the Service Request
Enable Register.
Service Request Enable Query
Returns the integer value of the Service Request Enable Register, with
bit 6 set to 0.
Read Status Byte Query
Returns the integer value of the Status Byte, with bit 6 as the master
summary bit. Note that a serial poll returns bit 6 as the RQS message.
Trigger
When parsed, this command causes the instrument to request a Single
Scan. If a scan is already in progress when this command is parsed, an
additional scan is not executed.
Self Test Query
Causes an internal self test to be run, returning the result as an integer
(representing the binary encoded value.) The self test does not require
any local operator interaction and returns the instrument to the power-up
state. The binary coding is:
Bit
Binary Value
0
1
1
2
2
4
3
8
4
16
5
32
6
64
7
128
8
256
9
512
10
1024
11
2048
12
4096
Wait-to-continue
This command is accepted by Hydra, but has no effect. *WAI is required
by the IEEE-488.2 standard, but is non-operational in Hydra.
Using the Computer Interface
Computer Interface Command Set
Error
ROM test failed (bad CRS)
External RAM test failed
Internal RAM test failed
Display selft test failed
Display bad or not installed
Instrument configuration corrupted.
EEPROM instrument configuration bad
EEPROM calibration data bad
A/D bad or not installed
A/D ROM test failed
A/D RAM test failed
A/D self test failed
Memory RAM test failed (2625A only).