2620A/2625A
Users Manual
Input Impedance
1 M in parallel with 100 pF maximum Maximum
Maximum Crest Factor
3.0
2.0 for rated accuracy
Crest Factor Error
Non-sinusoidal input signals with crest factors between 2 and 3 and pulse widths 100 µs
and longer add 0.2% to the accuracy specifications.
Common Mode Rejection
80 dB minimum at 50 or 60 Hz ± 0.1%, 1 k imbalance, slow rate
Maximum AC Input
300V rms or 424V peak on channels 0, 1, and 11
150V rms or 212V peak on channels 2 to 10 and 12 to 20
Voltage ratings between channels must not be exceeded
2 x 10
1 x 10
DC Component Error
SCAN and first MONitor measurements will be incorrect if the dc signal component
exceeds 60 counts in slow rate or 10 counts in fast rate. To measure ac with a dc
component present, MONitor the input and wait 5 seconds before recording the
measurement.
Cross-Talk Rejection
Refer to Appendix D.
A-8
Frequency
20 Hz - 50 Hz
50 Hz - 100 Hz
100 Hz - 10 kHz
10 kHz - 20 kHz
20 kHz - 50 kHz
50 kHz - 100 kHz
6
Volt-Hertz product on any range, normal mode input
6
Volt-Hertz product on any range, common mode input
Maximum Input at Upper Frequency
300 V rms
300 V rms
200 V rms
100 V rms
40 V rms
20 V rms