Fluke 2620A User Manual page 157

Hydra series ii data acquisition unit, hydra series ii data logger
Hide thumbs Also See for 2620A:
Table of Contents

Advertisement

Input Impedance
100 M minimum in parallel with 150 pF maximum
Common Mode and Normal Mode Rejection
See Specifications, DC Voltage Inputs
Cross-Talk Rejection
Refer to Appendix D
Open Thermocouple Detect
Small ac signal injection and detection scheme before each measurement detects greater
than 1 to 4 k as open. Performed on each channel unless defeated by computer
command.
RTD Inputs
Type
DIN/IEC 751, 100e Platinum (385)
RTD
Temperature
Resolution
(°C)
Slow
-200.00
0.02
0.00
0.02
100.00
0.02
300.00
0.02
600.00
0.02
* Sensor inaccuracies are not included
RTD
Temperature
Resolution
(°C)
Slow
-200.00
0.02
0.00
0.02
100.00
0.02
300.00
0.02
600.00
0.02
* Sensor inaccuracies are not included
IEC 751, Amendment 1, 100 e Platinum (IPTS-68)
4 Wire Accuracy* (±°C)
18°C to 28°C
Fast
90 Day
Slow
0.1
0.06
0.1
0.09
0.1
0.10
0.1
0.14
0.1
0.19
IEC 751, Amendment 1, 100 e Platinum (ITS-90)
4 Wire Accuracy* (±°C)
18°C to 28°C
Fast
90 Day
Slow
0.1
0.06
0.1
0.09
0.1
0.13
0.1
0.17
0.1
0.30
1 Year
1 Year
Slow
Fast
0.06
0.48
0.09
0.55
0.11
0.58
0.14
0.65
0.20
0.76
1 Year
1 Year
1 Year
Slow
Fast
Slow
0.06
0.48
0.09
0.55
0.13
0.60
0.18
0.69
0.31
0.87
Appendices
Specifications
0°C to 60°C
1 Year
1 Year
Slow
Fast
0.07
0.49
0.13
0.59
0.17
0.64
0.24
0.75
0.37
0.93
0°C to 60°C
1 Year
Fast
0.07
0.49
0.13
0.59
0.19
0.67
0.28
0.79
0.48
1.04
A
A-5

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

2625a

Table of Contents