Calibration; Items Of The Calibration Mode - A&D SW-D Setup Instructions

Weighing indicator for super washdown scales
Table of Contents

Advertisement

When calibrating, refer to this set up instructions. (The contents of SW instruction manual
are not the same as this set up instructions.)

7.1. Items of the calibration mode

The calibration mode has the following three functions.
• Setting of the weighing capacity, minimum division, and decimal point position.
• Gravity acceleration correction
• Calibration using a weight
How to enter the calibration mode
1. Make sure that the scale is in the weighing mode (displaying "kg", "g", "lb", "oz" or
"pcs").
2. Remove the screw protection caps on the load cell panel on the bottom of the
display. Loosen the four screws and then open the load cell panel. The calibration
(CAL) switch is located inside.
3. Press the CAL switch. The indicator displays "
or "7.4".)

7. CALIBRATION

Calibration (CAL) switch
10
Cal5et
". (And proceed to "7.2", "7.3",
Load cell panel