Calibration; Calibration Mode - A&D SJ-3000WP-BT Instruction Manual

Ip67 compact bench scale
Table of Contents

Advertisement

Adjusts the scale for accurate weighing. Calibrate the scale in the following cases.
When the scale is first installed.
When the scale has been moved.
When the ambient environment has changed.
For regular calibration.
Note: The Legal for Trade models can not be re-calibrated if they have been sealed.

11.1. Calibration Mode

The calibration mode has the following three functions.
Gravity acceleration correction
Calibration using a weight
Restoring the factory set values
How to enter the calibration mode
Method 1 :
1.
Make sure that the scale is in the weighing mode.
2.
Press and hold the
the switch.
Note: The above operation is disabled for the Legal for Trade models.
Method 2 :
1.
Make sure that the scale is in the weighing mode.
2.
Loosen the two screws on the switch cover and
open the switch cover. The calibration (CAL)
switch is located inside.
3.
Press the
CAL

11. CALIBRATION

switch until the "CAL" appears and release
SAMPLE
switch. The scale displays the "CAL".
S w i tc h c ov e r
C A L s w i tc h
28