7. Place the calibration weight with the same value
as displayed on the weighing pan, and wait for
the STABLE indicator to turn on.
8. Press the TARE key. The scale calibrates SPAN and
end
"
" is displayed. After displaying "
complete.
9. To finish the setting procedure, press the
the CAL switch. The indicator turns the power off.
Note: If the scale will be moved to another location, set the gravity acceleration
value for the new location and calibrate the scale according to the procedure
above. Refer to the previous section to set the gravity acceleration value.