Owon SDS series User Manual page 70

Sds series smart digital storage oscilloscopes
Hide thumbs Also See for SDS series:
Table of Contents

Advertisement

The "Automatic Measurements" menu is described as the following table:
Function Menu
Type
Source
Add
Show all
Add
Remove all
Remove
Type
Remove
Measure
The measured values can be detected on each channel simultaneously. Only if the
waveform channel is in the ON state, the measurement can be performed. The
automatic measurement cannot be performed in the following situation: 1) On the
saved waveform. 2) On the mathematical waveform. 3) On the XY format. 4) On
the Scan format.
Measure the frequency, the peak-to-peak voltage of the Channel CH1 and the mean,
the RMS of the Channel CH2, following below steps:
1. Press the Measure button to show the automatic measurement function menu.
2. Press the H1 button to display the Add menu.
3. Press the F2 button and choose CH1 as the source.
4. Press the F1 button, the type items will display at the left of screen, and turn the
5.Advanced User Guidebook
Figure 5-56 Measure Menu
Setting
Description
Press F1 ,show the measure types
CH1
Select the source
CH2
Show all the measures on the screen
Add the selected measure types (shown
at the left bottom, you could only add 8
types at most)
Remove all the measures
Turn the M knob, select the types need
to be deleted.
Remove the chosen measure type
65