Advertisement

5.Advanced User Guidebook
Operation steps:
1. Press the Display button and call out the Display Set Menu.
2. Press the H3 menu selection button to set XY Mode ON. The display format is
changed to be XY mode (see Figure 5-35).
Figure 5-35 XY Display Mode
Cymometer
It is a 6-digit cymometer. The cymometer can measure frequencies from 2Hz to the
full bandwidth. Only if the measured channel has triggering signal and in Edge mode,
it can measure frequency correctly. In the Single trigger mode, it is a one channel
cymometer and it can only measure the frequency of the triggering channel. In the
ALT trigger mode, it is a two channel cymometer and it can measure the frequency of
two channels. The cymometer is displayed at the right bottom of the screen.
To turn the cymometer on or off:
1. Press the Display button.
2. In the Display menu, press the H4 button to toggle between the cymometer
display ON or OFF.
VGA Output (only for the model with "V")
The VGA port could be connected to a computer monitor. The image of the
oscilloscope can be clearly displayed on the monitor.
To set the VGA Output:
1. Press the Display button.
2. In the Display menu, press the H5 button to toggle between ON or OFF.