Acquiring The Waveform; Making A Single Acquisition - Fluke 125 User Manual

Industrial scopemeter
Hide thumbs Also See for 125:
Table of Contents

Advertisement

Fluke 125
Users Manual

Acquiring the Waveform

Making a Single Acquisition

To catch single events, you can perform a single shot.
(One time screen update.) To set up the test tool for a
single shot on the input A waveform, do the following:
Connect the probe to the signal to be measured.
Open the application menu.
Open the TRIGGER... submenu.
Highlight A.
Accept INPUT: A.
Highlight SINGLE.
Accept the Single shot setup.
(2x)
2-14
The test tool will now have a screen like Figure 2-13.
Wait: appears on bottom of the screen to indicate that the
test tool is waiting for a trigger.
Run: appears on bottom of the screen when the single
acquisition is triggered.
Hold: appears on bottom of the screen when the single
acquisition has been completed.
Figure 2-13. Making a Single Acquisition
To perform a next single acquisition, do the following:
Wait for another single acquisition
trigger.

Advertisement

Table of Contents
loading

Table of Contents