Fluke 125 User Manual page 101

Industrial scopemeter
Hide thumbs Also See for 125:
Table of Contents

Advertisement

Input A
Ohm (Ω)
Ranges......................50Ω, 500Ω, 5 kΩ, 50 kΩ, 500 kΩ,
............................................................. 5 MΩ, 30 MΩ
Accuracy: ......................................... ±(0.6% +5 counts)
.........................................50 Ω ±(2 % + 20 counts)
Full Scale Reading:
50 Ω to 5 MΩ........................................... 5000 counts
30 MΩ ..................................................... 3000 counts
Measurement Current ..........................0.5 mA to 50 nA
decreases with increasing ranges
Open Circuit Voltage ............................................... <4V
Continuity (CONT)
Beep..................................... <(30Ω ±5Ω) in 50 Ω range
Measurement Current ........................................ 0.5 mA
Detection of shorts of ..........................................≥1 ms
Diode
Measurement Voltage:
@0.5 mA ...........................................................>2.8V
@open circuit........................................................<4V
Accuracy.............................................. ±(2% +5 counts)
Measurement Current.........................................0.5 mA
Polarity ..................................... + on input A, - on COM
Capacitance (CAP)
Ranges ................. 50 nF, 500 nF, 5 μF, 50 μF, 500 μF
Accuracy............................................ ±(2% +10 counts)
Full Scale Reading .....................................5000 counts
Measurement Current............................ 5 μA to 0.5 mA
increases with increasing ranges
Dual slope integrating measurement with parasitic
serial and parallel resistance cancellation.
10
Specifications
Dual Input Meter
10-7