Function List - A&D SW-150KL Instruction Manual

Sw series super wash down scales
Hide thumbs Also See for SW-150KL:
Table of Contents

Advertisement

12.3. Function list

Item
Auto power-off function
poff
Weight display
resolution
re5o
Weighing unit
Ut-g
Zero tracking
trc
Stability band
width
5t-b
Stability
detection time
5t-t
Stability
Response speed
Cond
Serial interface
baud rate
bp5
Serial interface
data bits / parity
btpr
Serial interface
Data output
mode
prt
Serial interface
Type
5if
Serial interface
Address
adr
Serial interface
Response
aCk
Comparator mode
Cp-l
Parameter
0
Auto power-off disabled
1
Auto power-off enabled
0
Normal (1/3,000)
1
High (1/6,000 or 1/7,500)
2
Maximum (1/12,000 or 1/15,0 00)
0
kg
1
g
0
Zero tracking function disabled
1
Zero tracking function enabled
0
±0.5d (width 1d)
1
±1.0d (width 2d)
2
±2.0d (width 4d)
0
0.5 second
1
1.0 second
2
2.0 seconds
0
Weak stability / fast response
1
Normal stability / normal response
2
Strong stability / slow response
0
2400 bps
1
4800 bps
2
9600 bps
0
7 bits / EVEN
1
7 bits / ODD
2
8 bits / Non parity
0
Stream mode / Command mode
1
Command mode only
2
Print switch mode /
3
Auto-print mode +/- data / Command mode
4
Auto-print mode + data/ Command mode
0
RS-232C
1
RS-422
2
RS-485
## = 00 to 99
##
(Factory setting: ## = 01)
0
No reply except data to commands
1
Reply to commands
0
Five-level (LOLO, LO, OK, HI, HIHI)
1
Three-level (LO, OK, HI)
Description
Command mode
23
Turns the power OFF
automatically
Changes the
minimum display
Tracks the zero drift
Conditions to turn the
STABLE indicator on
d = minimum display
in kg
Filtering
Only when the
options, RS-232C or
RS-422/485, are
used.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sw-150kmSw-15ksSw-60klSw-6ksSw-15kmSw-30km ... Show all