The installation procedure is the same as for OP-03. Refer to "13.2.1. Installing OP-03".
Set the function settings "
necessary.
Before using OP-04, the function setting "
422 or RS-485 is used.
To connect more than one scale to a computer, set a different address to each
scale using the function setting "
13.3.2. OP-04 Specifications
RS-422/485 Specifications
Transmission system EIA RS-422 / 485
Transmission form
Data format
Circuit diagram
bp5
btpr
", "
adr
Asynchronous, bi-directional, half-duplex
Baud rate:
2400, 4800, 9600 bps
Data bits:
7 bits + parity 1bit (EVEN / ODD)
or 8 bits (non parity)
Start bit:
1 bit
Stop bit:
1 bit
Code:
ASCII
Terminator: C
L
R
F
LSB
0
1
Start bit
100Ω
prt
5if
", "
", "
", "
5if
" must be set to specify whether RS-
".
(C
:0Dh, L
:0Ah)
R
F
MSB
2
3
4
5
6
Data bits
Parity bit
1 SDA (RS-422 Out)
2 SDB (RS-422 Out)
3 RDA (RS-422 In / RS-485 In-Out)
4 RDB (RS-422 In / RS-485 In-Out)
5 TRM (Termination)
6 SG (Signal ground)
7 SHLD (Shield)
1 SDA (RS-422 Out)
2 SDB (RS-422 Out)
3 RDA (RS-422 In / RS-485 In-Out)
4 RDB (RS-422 In / RS-485 In-Out)
5 TRM (Termination)
6 SG (Signal ground)
7 SHLD (Shield)
35
adr
aCk
" and "
" as
1 (-2V~-5V)
0 (2V~15V)
Stop bit
A-B