The scale has function settings to specify the scale performance.
The parameters set in the function settings are maintained even if the power is turned
OFF.
Item

12.1. Setting the parameters

1. Turn the power OFF.
2. Press and hold the TARE switch and press the
ON/OFF switch to turn the power ON. The software
version is displayed.
"∗∗∗" indicates the software version number.
3. After about one second, the item is displayed.
4. Change the item or parameter using the following switches.
UNITS
PRINT
5. Press the
end
After "
", the power is turned OFF automatically.
To cancel the setting procedure without storing the value, press
the ZERO switch. The power is turned OFF automatically.
12.2. To restore the function settings to the
factory setting values
1. Turn the power OFF.
2. Press and hold the TARE switch and press the
ON/OFF switch to turn the power ON. Only release the
ON/OFF switch and hold the TARE switch until
Clrf no
"
" with "
3. Press the PRINT switch.
Clrf no
"
" changes to "
Clrf go
4. When "
The factory setting values are restored. After "
is turned OFF automatically.
To cancel the restoring procedure, press the ZERO switch.
The power is turned OFF automatically..

12. FUNCTIONS

To display the next item.
To increase the value of the blinking
digit by one (to change the parameter).
When the parameter is changed, the
STABLE indicator turns off.
TARE
switch to store the setting value.
no
" blinking is displayed.
Clrf go
" is displayed, press the TARE switch.
Parameter
go
" with "
" blinking.
end
", the power
22
While
holding down,
press
Press
When changed
Press
Press
Press