4-1 Smart Range Function

The GX-8202MD, GF-8202MD, GX-32001MD and GF-32001MD are equipped with two ranges of
"precision range" of a higher resolution and "standard range" of normal resolution.
The range is switched automatically depending on the value displayed.
Placing a heavy container on the weighing pan and pressing the
balance to weigh in the precision range. (Smart range function)
The range can be fixed to the standard range, by pressing the SAMPLE key.
Note
Once the range is switched to the standard range, it will not switch to the precision range
automatically even when the displayed value becomes within the precision range value. Press
the RE-ZERO or SAMPLE key to use the precision range again.
Example
GX-32001MD and GF-32001MD, precision range = 6.2 kg x 0.1g,
standard range = 32.2 kg x 1 g.
Step
1 Press the RE-ZERO key.
The balance will start weighing, using the
precision range.
Step
2 Place a container on the weighing pan.
When the weighing value exceeds the
precision
switched to the standard range.
Step
3 Press the RE-ZERO key.
The balance will be switched to the precision
range.
Step
4 Place a sample on the pan.
When the weighing value is within the
precision range, the balance will perform a
weighing using the precision range.
Precision range/standard range value
GX-32001MD
GF-32001MD
GX-8202MD
GF-8202MD
range,
the
range
Precision range
(after RE-ZERO key is pressed)
g
Up to 6200.9
g
Up to 2200.09
will
be
g
g
2200.1
16
RE-ZERO
Weighing Pan
Precision range
Tare (Container)
Standard range
Precision range
Sample
Precision range
Standard range
6201
g to 32208
g to 8200.8
key allows the
g
g