Calibration Instructions for the Type
Approval Version of the SJ series

When is to Calibration Required?

Calibration may be required when the SJ series initially installed or if the scale is moved a
substantial distance.

Calibration using a Calibration Weight

1. Entering calibration mode.

Remove the calibration switch cover at the bottom of the device
and press the calibration switch while the device is in the
Weighing Mode.
Cal
Then
will be displayed.

2. Zero calibration

Press the [RE-ZERO] key to calibrate zero.
Wait for the Stable Indicator to be displayed then press the [RE-ZERO] key.
Calf
will be displayed after a few seconds. Press the [BLIND]* key to perform only a
Zero Calibration. The scale will then automatically return to the Weighing Mode.

3. Span calibration

Calf
When
is displayed, place the weight at the center of the platform. Wait for the
Stable Indicator to be displayed, then press the [RE-ZERO] key. The display will show
end
and the scale will automatically return to the Weighing Mode.
See "SPECIFICATIONS" about the calibration weight.

Calibration by Gravity Compensation

If the acceleration of gravity at your location is not 9.798 m/s
calibration weights, the scale can be calibrated by compensating for the acceleration of
gravity. (Refer to "The Value of Gravity at Various Locations".)

1. Setting a new acceleration value.

Cal
When
is displayed, press the [BLIND]* key and
the [RE-ZERO] key to increment the blinking digit and press the [BLIND]* key to move the
blinking digit.

2. Storing the value in the memory.

While pressing the [BLIND]* key, press and hold the [RE-ZERO] key and release the
[BLIND]* key. Then the display will show
off to finish the procedure.
Remarks*: The [BLIND] key can be activated by pressing the area between the [ON/OFF] and
[RE-ZERO] keys.
Calibration switch cover
Cal0
will be displayed.
9. 7 98
end
and return to
1/2
1WMPD4001897
Calibration switch
2
and you do not have
will be displayed. Press
Cal
. Turn the scale

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SJ-12KH and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for A&D SJ-12KH

  • Page 1 1WMPD4001897 Calibration Instructions for the Type Approval Version of the SJ series When is to Calibration Required? Calibration may be required when the SJ series initially installed or if the scale is moved a substantial distance. Calibration switch Calibration using a Calibration Weight 1.
  • Page 2: Specifications

    SPECIFICATIONS MODEL SJ-1000 SJ-2000 SJ-5000 SJ-12K 0 ~ 10 kg x 10 g Capacity/Resolution 1000 g x 1 g 2000 g x 2 g 5000 g x 5 g 10 ~ 12 kg x 20 g Maximum tare 1000 g 2000 g 5000 g 9.99 kg...

This manual is also suitable for:

Sj-2000hSj-5000hSj-1000hSj-1000Sj-2000Sj-5000 ... Show all

Table of Contents