9. Function Switch and Initialization

Permit or Inhibit

9.1.
The balance stores parameters that must not be changed unintentionally.
Example: Calibration data for accurate weighing, Data for adapting to the operating environment,
Control data for the RS-232C interface.
The balance is equipped with five switches for the purpose of protecting parameters as shown above.
Each switch can select either "permit" or "inhibit". "Inhibit" protects parameters against
unintentional operations.
In addition to the above, when using balances with software version 1.660 or later, the password
function can be used to restrict calibration functions.
1 Press the
ON:OFF
2 While pressing and holding the
press the
ON:OFF
3 Press the
PRINT
4 Specify the switches using the following keys.
key ........... To select the blinking digit.
RANGE
key ........ To change the parameter of the selected switch.
RE-ZERO
key ............ To store the new parameter and return to the weighing mode.
PRINT
key ................ To cancel the operation ( Clr is displayed ). To return to the weighing
CAL
Function table
Calibration using the internal mass
Calibration using an external weight
Automatic self calibration (for variation of ambient temperature)
Internal mass correction
key to turn the display off.
PRINT
key to display
p5
key. Then the balance displays the function switches.
0
To inhibit changes. (Cannot be used.)
1
To permit changes. (Can be used.)
mode, press the
(The display shown left indicates the factory settings)
0
To inhibit changes to the function table.
1
To permit changes to the function table.
0
Software ver. 1.00 to 1.64
Software ver. 1.660 or later pA55 0
1
To permit calibration using the internal mass.
0
Software ver. 1.00 to 1.64
Software ver. 1.660 or later pA55 0
1
To permit calibration using an external weight.
0
To inhibit automatic self calibration.
1
To permit automatic self calibration.
0
To inhibit correction.
1
To permit correction.
Note 1: The operation is permitted when logged in as an administrator ( Adm
key and the
RANGE
.
key once again.
CAL
To inhibit calibration using the internal mass.
To inhibit calibration using the internal mass.
pA55 1 , 2 To inhibit calibration using the internal mass when
logged in as a user (U5
See Note 1.
To inhibit calibration using an external weight.
To inhibit calibration using an external weight.
pA55 1 , 2 To inhibit calibration using an external weight when
logged in as a user (U5
See Note 1.
44
key,
) or a guest (GUE
ER
) or a guest (GUE
ER
).
5T
).
5T
).
IN