Table 1-1
VERTICAL AMPLIFIER
Table 1-2
HORIZONTAL AMPLIFIER
Characteristics
Input Signal Amplitude (Differential)
Bandwidth (6-Division Reference)
Risetime (5-Division Reference)
Aberrations (6-Division Reference)
Position Effect on Aberrations
(6-Division Reference with a
067-0680-00 Calibration Fixture)
Vertical Centering
Delay Line Length
Modes
Rate
Chop
- .
Alt
Bandwidth
Performance Requirements
Dc to at least 85 MHz with a 067-0680-00
Calibration Fixture.
Dc to at least 50 MHz with a calibrated
5A48.
4 ns or less with a 067-0680-00 Calibration
Fixture.
7 ns or less with a calibrated 5A48.
5% or less measured with a 067-0680-00
Calibration Fixture.
4% or less measured with a calibrated 5A48.
Chop and alt.
Once every two sweeps.
Characteristics
1
DC to at least 2 MHz.
Supplemental Information
50 mvldiv,
2
2%. Less than 0.5%
difference between left and right vetical
plug-in compartments.
Front corner aberrations of +step or
-step response signal should not
exceed
k
5% when the waveform is
positioned not more than 1 div beyond
graticule center.
Within
k
0.5 div of graticule center.
I
140 ns
50 kHz +5O0/0 -30%; 3 ps on, 2 ps
off.
Eight-division signal used as a
reference.
Performance Requirements
Supplemental Information
I
20 kHz.
Horizontal Centering
X-Y Operation
REV MAY 1982
Less than 2" phase shift from dc to at least
Within 0.5 div of graticule center.