Chapter 8: Troubleshooting
PROBLEM
Fume hood blower
operates but lights
don't work (cont.)
Contaminants
escape outside of
fume hood
Vertical sash no
longer operates
smoothly
Combination A-
Style sash no longer
operates smoothly
46
CAUSE
CORRECTIVE ACTION
Lamp wiring is
Inspect lamp wiring.
disconnected.
Defective lamp
Replace lamp ballasts.
ballasts.
Improper user
See "Certifying the Hood" Chapter 3
techniques for the
and "Safety Precautions" Chapter 4
fume hood.
sections in the manual. (Ref. Appendix
D)
Restriction of the
Remove baffles to ensure that all air
baffle air slots or
slots and the exhaust outlet are
blockage of the
unobstructed.
exhaust outlet.
External factors are
See "Location Requirements" Chapter
disrupting the fume
2, "Certifying the Hood" Chapter 3,
hood airflow patterns
and "Safety Precautions" Chapter 4
or acting as a source
sections of this manual. (Ref. Appendix
of contamination.
D)
Fume hood has
Have fume hood re-certified and check
improper face
remote blower exhaust system. Hood
velocity.
should have average face velocity of
80-100 fpm.
Cable is frayed or
Inspect cable and replace cable if worn
plastic protection is
or damaged immediately; otherwise
damaged.
injury could result.
Pulley bearing is
Replace pulley, bearing or add grease.
damaged.
Cable has slipped off
Re-install; cable must be replaced
the pulleys.
immediately if damaged.
Weight has broken
Replace weight pulleys.
pulleys.
Horizontal glass
Re-install horizontal glass on tracks.
panels have come off
the tracks.
Product Service 1-800-522-7658