Operating the Vertical-
Rising Sash
Because of the Protector Radioisotope Hood's
counterbalanced sash mechanism, it will take only a
few pounds of force to move the sash up or down, and
you can operate the sash smoothly with one or two
hands positioned along the handle. The vertical-rising
sash may be raised to a maximum 28" operating height.
The airflow requirements should be sized for the 28"
operating height; if using sash stops then the airflow
requirements can be reduced by approximately 40% at
18" or approximately 50% at 15".