Chapter 2: Prerequisites
Service Line Requirements
All service lines to the laboratory fume hood should be
¼ inch outside diameter, copper (brass for natural gas),
and equipped with an easily accessible shut-off valve,
should disconnection be required. If the service line
pressure exceeds 40 PSI, it must be equipped with a
pressure regulator to reduce the line pressure.
Space Requirements
The dimensions for the different models are shown in
Appendix B: Protector Radioisotope Hood Dimensions.
Your Next Step
After you have determined that the location you have
selected accommodates the installation and operational
requirements of your fume hood, you are ready to begin
installation. Proceed to Chapter 3: Getting Started.