7 Troubleshooting
7.1 Fault actions
Component
Problem
UV measure-
Low sensitivity
ment
Error in external chart
recorder
pH measurement
Component
Problem
pH measure-
No response to pH
ment
changes
Small response to pH
changes
Slow pH response or cali-
bration impossible
pH values vary with varied
back-pressure
98
Corrective action
Aging lamp. Check the lamp and replace if neces-
sary
Wrong lamp position. Check that the lamp position
and the wavelength used (filter position) fit together.
Dirty on-line filter. Clean or replace the filter.
Check the chart recorder in accordance with its
manual.
Test the recorder function by selecting recorder
test according to Check recorder, on page 68.
Corrective action
Check that the electrode cable is connected prop-
erly to the rear of the module.
The electrode glass membrane may be cracked. If
so, replace the electrode.
Clean the pH electrode according to Section 6.3.4
Clean the pH electrode, on page 89 and recalibrate.
If the problem remains, replace the pH electrode.
Check the electrode glass membrane. If it is contam-
inated, clean the electrode following the instructions
in Section 6.3.4 Clean the pH electrode, on page 89.
If the membrane has dried out, the electrode may
be restored by soaking it in buffer overnight.
Clogged liquid junction. Refer to Section 6.3.4 Clean
the pH electrode, on page 89.
Replace the pH electrode.