GE Monitor UPC-900 Operating Instructions Manual page 75

Advertisement

Step
8
Conductivity graph
The graph shows the conductivity of 1.00 M NaCl at 20 to 30°C.
Monitor UPC-900 Operating Instructions 29-0567-16 AB
Action
The current value is displayed as default. Enter the
theoretical conductivity value according to the graph
and press OK. The new cell constant is automatically
calculated. The range is 1.000 to 999.9 mS/cm.
97
95
90
85
80
77
20
25
30
5 Operation
5.10 Monitor setup
Warning! This will
change cell calib
Display
Setup Adjust Cond
(83.53mS/cm)
86.6
75