7 Troubleshooting
7.1 Fault actions
Conductivity measurement
Component
Problem
Conductivity
Incorrect or unstable
measurement
reading
Baseline drift or noisy sig-
nal
Conductivity measure-
ment with the same
buffer appears over time
100
Corrective action
Check that the conductivity flow cell cable is con-
nected properly to the reading rear of the module.
Check that the pump operates correctly.
If temperature compensation is being used, check
that the temperature sensor is calibrated, and that
the correct temperature compensation factor is in
use.
Check that the column is equilibrated. If necessary,
clean the column.
Check the operation of the mixer.
There may be air in the flow cell. Use a flow restric-
tor after the flow cell and check that the flow restric-
tor gives a back-pressure of 0.2 MPa.
Check for leaking tubing connections.
Check that the column is equilibrated. If necessary,
clean the column.
Check the operation of the mixer and the pump.
Clean the flow cell according to the procedure in
Section 6.3.2 Clean UV flow cell, on page 86.
Clean the flow cell according to the procedure in
Section 6.3.2 Clean UV flow cell, on page 86.
The ambient temperature may have changed. Use
a temperature compensation factor, see Setup
conductivity temperature compensation, on page72.