Drying in the Cabinet
Start-up
• Turn off UV light and turn the UV Security switch to "OFF". Remove the key and
store it in a safe place.
• Ensure the drain strainer is in place.
• Turn the switch to "Fan".
• Check the condition of the prefilters. If your protocol requires the use of evidence
prefilters, ensure a new set of both door and cabinet prefilters has been properly
installed, according to the instructions in
Precautions.
• Wash hands and arms thoroughly with germicidal soap.
• Wear a long-sleeved lab coat with knit cuffs and over-the-cuff rubber gloves. Use
protective eyewear. Wear a protective mask if appropriate.
Wipe-Down
• Open the door to its fully open position.
• Wipe down the interior surfaces of the cabinet with 70% ethanol, or a suitable
disinfectant, and allow it to dry.
CAUTION: Do not use bleach on the stainless steel liner. Use disinfectants
compatible with stainless steel.
Loading Evidence
• Load the evidence so that air flows around all surfaces. Do not overload the
cabinet. Do not place objects directly against the side walls.
• Do not obstruct the intake prefilter (bottom of the door(s)) or the exhaust air
discharge, on top of the cabinet.
• Large objects should not be placed close together.
• Slowly close the door(s) and use a padlock (not supplied) or secure it with a
chain of custody seal.
Note if the Filter System light is on.