Do not operate the cabinet without all of the appropriate filters in place. If
an odor is detected coming out of the granular activated carbon, it needs to
be replaced.
The weight limit for the fold down shelves is 15 lbs (6.8 kg) each, and the
floor of the cabinet and lower storage area is 20 lbs (9.1 kg). The weight
limit for the hanging rod is 20 lbs. (9.1 kg).
When surface disinfecting the cabinet:
• Always wear appropriate personnel protective equipment.
• NEVER use high pressure water to clean the liner; this practice
could generate biohazardous aerosols.
• NEVER spray disinfectants into the blower intake at the top of the
liner; this may damage the blower.
• Avoid splashing the disinfecting solution on skin or clothing.
• Ensure adequate room ventilation.
• Carefully follow the disinfectant's safety instructions.
• Always dispose of disinfecting solutions in accordance with local and
national laws.
• DO NOT allow bleach or disinfectants with high concentrations
of free chlorine to contact the stainless steel components of
the cabinet for a long period of time. Free chlorine will corrode
stainless steel after extended contact.
When handling used filters or prefilters, always wear appropriate personnel
protective equipment.