Electrical Requirements
The product models have the following electrical requirements.
Catalog Number
33900-00
33900-01
*Electrical Requirements, 'V' = VAC (Voltage with alternating current), 'A' = Amperes
A dedicated outlet with an appropriate circuit breaker should be located as close
as possible to the product (6 feet or 1.8 meters maximum). Consult your local
electrical codes for properly rated circuit breakers. For safe operation, the
dedicated outlet must provide a protective earthing ground connection to the
product.
Exhaust Requirements
If you intend to connect the cabinet to the optional Exhaust Connection Kit, first examine
the location to ensure that it accommodates the cabinet's exhaust duct kit. The area
directly above the cabinet's exhaust should be clear of structural elements, water and
utility lines, or other fixed obstructions. There should be enough clearance to
accommodate a 5-inch diameter duct. For further information about the cabinet's
exhaust system requirements, please refer to
Typical Operating Current
(Amps)
6A
3A
Appendix D:
10
Electrical Circuit Requirements
115 VAC, 60 Hz, 6A
230 VAC, 50/60 Hz, 3A
Accessories.
*
1 Phase
1 Phase