Cable information:
Cable Name
Adapter Cable
Important information regarding Electro Magnetic Compatibility (EMC)
This electrical medical equipment needs special precautions regarding EMC and put
into service according to the EMC information provided in the user manual; The
equipment conforms to this IEC 60601-1-2:2020 standard for both immunity and
emissions. Nevertheless, special precautions need to be observed:
1. Use of accessories and cables other than those specified or provided by the
manufacturer of Fast-Pack pro could result in increased electromagnetic emissions
or decreased electromagnetic immunity of Fast-Pack pro and result in improper
operation.
2. Use of Fast-Pack pro adjacent to or stacked with other equipment should be
avoided because it could result in improper operation. If such use is necessary,
Fast-Pack pro and the other equipment should be observed to verify that they are
operating normally.
3. Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to
any part of the Ultra X, including cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment could result.
4. If the use location is near (e.g. less than 1.5 km from) AM, FM or TV broadcast
antennas, before using this equipment, it should be observed to verify that it is
operating normally to assure that the equipment remains safe with regard to
electromagnetic disturbances throughout the expected service life.
Cable Length (m)
2
34 / 36
Shielded or not
No
Remark
/
Need help?