Pre-
Cleaning:
Cleaning:
Disinfection:
Do a manual pre-cleaning, until the components are visually clean.
Submerge the components in a cleaning solution and flush the
lumens with a water jet pistol with cold tap water for at least 10
seconds. Clean the surfaces with a soft bristol brush.
Regarding cleaning/disinfection, rinsing and drying, it is to
distinguish between manual and automated reprocessing
methods. Preference is to be given to automated reprocessing
methods, especially due to the better standardizing potential and
industrial safety.
Automated Cleaning:
Carefully put the components into the washer-disinfector on a tray
and set the parameters as follows, then start the program:
•
4 min pre-washing with cold water (<40°C);
•
emptying
•
5 min washing with a mild alkaline cleaner at 55°C;
•
Emptying
•
3 min neutralising with warm water (>40°C);
•
emptying
•
5 min intermediate rinsing with warm water (>40°C);
•
Emptying
The automated cleaning processes have been validated by using
0.5% neodisher MediClean forte (Dr. Weigert).
Note Acc. to EN ISO 17664 no manual reprocessing methods are
required for these devices. If a manual reprocessing method has
to be used, please validate it prior to use.
Use only approved washer-disinfectors according to EN ISO
15883, maintain and calibrate it regularly.
Follow instructions and observe concentrations given by the
manufacturer (see general recommendations).
Automated Thermal Disinfection in washer/disinfector under
consideration of national requirements in regards to A0 value (see
EN ISO 15883).
21 / 36
Need help?
Do you have a question about the Fast-Pack pro and is the answer not in the manual?
Questions and answers