Eighteeth E-connect S User Manual
Hide thumbs Also See for E-connect S:

Advertisement

0197
S
E-connect
USER MANUAL

Advertisement

Table of Contents
loading

Summary of Contents for Eighteeth E-connect S

  • Page 1 0197 E-connect USER MANUAL...
  • Page 2: Table Of Contents

    Content 1. Scope of E-connect S ........................4 Parts Identification ......................4 Components and Accessories ..................5 Options (sold separately) ....................5 2. Symbols used in the User Manual ....................6 3. Before Use ............................7 Intended Use ........................7 Contraindications ......................
  • Page 3 8.2.4 The water quality has to be convenient to the local regulations especially for the last rinsing step or with a washer-disinfector............... 32 8.2.5 Thoroughly clean and wash the components before autoclaving....32 8.2.6 Do not lubricate the motor handpiece..............32 8.2.7 Do not clean the contra angle with an ultrasonic cleaning device.
  • Page 4: Scope Of E-Connect S

    Scope of E-connect S 1. Scope of E-connect S 1.1 Parts Identification 1.Charge Base 2.Handpiece 3.Contra Angle 4.Insulating Sleeve 5.File Clip (2 pcs) 6.Lip Hook (2 pcs) 7.Measuring Wire 8.Adapter 9.Spray Nozzle Page 4 / 47...
  • Page 5: Components And Accessories

    Scope of E-connect S 1.2 Components and Accessories Motor Handpiece (1pcs) Charge Base (1pcs) Contra Angle (1pcs) Part No. 6051035 Part No. 6051036 Part No. 6036006 Adapter (1pcs) Measuring Wire (1pcs) Lip Hook (2pcs) Part No. 6016007+6016008 Part No. 6015011 Part No.
  • Page 6: Symbols Used In The User Manual

    Symbols used in the User Manual 2. Symbols used in the User Manual If the instructions are not followed properly, operation may WARNING lead to hazards for the product or the user/patient. Additional information, explanation of operation and NOTE performance. Serial number Catalogue number Manufacturer...
  • Page 7: Before Use

    HF Surgical Equipment in the hospital. Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the E-connect S, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
  • Page 8: Installing The E-Connect S

    Installing the E-connect S 4. Installing the E-connect S 4.1 Installation contra angle Make sure 4 pins on contra angle alignment the slots of handpiece, plug You can also use disposable sleeve them together until it “click” securely (sold separately) instead of insulating into place.
  • Page 9: Connecting Measuring Wire

    Installing the E-connect S WARNING Inspect the file head before inserting ● the file. Do not use the damaged file Single apex head. function only Make sure the motor is stopped when ● inserting and removing files. careful when inserting and ●...
  • Page 10 Installing the E-connect S If only need a base to put the device on Put the handpiece all the way into the dentist element of dental chair ( without charge base, the charge state will show charge function ), handpiece base is on the screen.
  • Page 11: Use Interface

    Interface 5.Use Interface 5.1 Panel key Turn Power On Press ● more than 0.5 seconds to turn on the instrument Memory Change Press < > during standby state Operation mode Change Press once during standby state, press < > to change, then press ●...
  • Page 12: Screen Display

    Interface 5.2 Screen display Standby interface ① Memory mode number ② Speed of this memory program ③ Unit of speed: Revolutions Per minute N cm · ④ Residual battery power ⑤ Operation mode ⑥ Torque of this memory program ⑦ Unit of torque: Newton centimeter Standard working mode ①...
  • Page 13 Interface Canal measurement standby interface ① Memory mode number M0 is apex stand-alone memory ② Electronic apex locator ③ Apex flash bar Canal measurement start interface ① Indication number The number has no represent of actual length, only for indication ②...
  • Page 14: Terms And Definition

    Interface 5.3 Terms and definition Forward ( Clockwise rotation ) Reverse ( Counter clockwise rotation ) Be applied to special file, inject calcium hydroxide and other solutions Reciprocation Be applied to reciprocating file, path file and rotary file protection by setting some special angle Adaptive torque control Up to setting torque, the motor will move with reciprocating mode;...
  • Page 15: Setting

    · standby state, press < > to change. E-connect S has 4 operation modes: FWD, REV, Op e ra ti on M od e REC and ATC (See chapter 5.3 Terms and definition to get the explanations of these modes).
  • Page 16 Be careful to use this function, very professional skill is needed, otherwise has risk of file broken. E-connect S integrated apex locator, if the lip hook is connecting with patient’s lip, when the endo file...
  • Page 17 Setting (without insulating sleeve) or file touch the patient’s lip or operator’s fingers (without insulating glove), take care to avoid this, the file rotated by motor has risk of injure someone. When the endo file out of root canal, the motor will Auto Stop not auto stop with default setting, Press <...
  • Page 18: Preset Programs

    Setting 6.3 Preset programs For convenience, we preset some common file system. Long press to entry preset program during F wd standby state, the interface will show as left. Protaper SX&S1 350 rpm ○ M1 ( ) meanings the current memory mode, you Protaper S2 3.
  • Page 19: Advanced Setting

    During power off state, holding down press then press ● to entry advanced setting, the version number software will appear on the display screen. E-connect S can update software very easy without Ve rsi on s tools and software. Contact your distributor to update if necessary.
  • Page 20 Setting Press again, the “Auto Return time” can be change, it means when setting parameters just like speed and torque, the system will back to standby Auto Return time interface if there is no operation in 5 seconds. 5 Se c to adjust, then press ●...
  • Page 21: Parameter Logic

    Setting Press again, entry “Restore setting” function, to select “ON”, press ● to start press < > recovering, all the parameters be set by operator will be recovered by default factory setting (See chapter 6.5 estore settings Parameter logic). NOTE After restore setting, all the parameters will be covered, record what you need before this operation.
  • Page 22 Setting The default advanced settings are listed below, the setting can be changed as needed. 10Min Auto Power off Startup memory Auto Return time 5Sec Language English Beeper Volume Calibration Habit hand Right hand Restore settings The speeds (rpm) in different operation mode are not the same, details are listed below.
  • Page 23 Setting The FWD Angle (degrees) and REV Angle (degrees) in different operation mode are not the same, details are listed below. 100 120 150 160 180 The same with the front table Angle 200 230 250 260 280 300 320 340 360 370 The same with the front The same with the front table table...
  • Page 24: Operation

    Operation 7.Operation 7.1 Charge Displays the present remaining amount of the battery. Less than 15% remains, please charge. NOTE If the power if less than 15%, must be recharged within 30 days, otherwise the battery will be damaged. If continue to use, the torque and speed will low LowPower than the setting value, and low power warming will Please C harge...
  • Page 25: Motor Operation

    5.2 Screen display). WARNING Use the E-connect S outside the oral cavity to make sure that the device is ● functioning properly. Change file on time to avoid file separation within the canal. File may separate ●...
  • Page 26 Operation Electromagnetic noise in surroundings environment may interfere with the ● device operation, do not rely on device’s automatic control completely, always pay attention to the feedback from display. NOTE If there is any abnormal functioning, stop using the device and report to company. The file separates more easily at high speeds, please follow the manufacturer’s recommendations of the speed and check the settings of the Endo motor before use.
  • Page 27: Apex Operation And Not Suitable Condition

    Operation NOTE The will not be able to perform a precise measurement for every time, especially in cases of abnormal or unusual morphology of the root canal. The user need coordinate with x-ray to check the results of the measurement. If the meter does not move when you enter the file, it is possible that the unit is not working normally, therefore, stop using.
  • Page 28 Operation NOTE The file clip must hold the file correctly. Push the button of file clip in the direction shown by the arrow, clip the holder onto the metal upper part of the file and then release the button. The file clip must be nearly vertical with file handle, otherwise will damage the tip of the file holder.
  • Page 29 Operation Root canal blood overflow from the opening If blood spills from the root opening and contacts the gums, it will cause leakage of electricity, which cannot be accurately Clean measured. Wait for the bleeding to stop completely. Clean the root canal and the opening, completely empty the root canal blood, and then measure it.
  • Page 30 Operation Retreatment canal which was filled with gutta-percha The gutta-percha must be completely removed to eliminate its insulation, then pass a small file all the way through the apical foramen and then put a little Gutta-percha saline in the canal, but do not let it overflow the canal opening.
  • Page 31 Operation Extremely dry canal If the canal is extremely dry, the meter may not work until it is quite close to the apex. In this case, try to moisten the canal with oxydol or Too dry saline. Difference measuring result between Apex locator...
  • Page 32: Cleaning, Disinfection And Sterilization

    Cleaning , Disinfection and Sterilization 8.Cleaning, Disinfection and Sterilization 8.1 Foreword For hygiene and sanitary safety purpose, the components (contra angle, file clip, lip hook and insulating sleeve) must be cleaned, disinfected and sterilized before each usage to prevent any contamination. This concerns the first use as well as the subsequent uses.
  • Page 33 Remove and disconnect the components (Contra Angle, Lip Hook, File clip, Insulating Sleeve) before cleaning. Refer to "Chapter 4-Installing E-connect S" of this manual for disassembly Preparation instructions. before cleaning WARNING Do not fail to take out the file before cleaning the ●...
  • Page 34 Cleaning , Disinfection and Sterilization Pre-cleaning: water temperature <30℃, 2 min; Cleaning: water temperature 45℃, 5 min; use an enzyme detergent solution (mild and aldehyde free solution) which is suitable to be used with washer-disinfector, and use in accordance with the IFU of the detergent solution manufacturer;...
  • Page 35 Cleaning , Disinfection and Sterilization to 141℃(286°F) and in accordance with EN ISO 11607. Sterilization Steam sterilization at 134℃, at least 6 minutes. Minimum drying time after sterilization: 10 minutes. WARNING Use only approved autoclave devices according ● to EN 13060 or EN 285. Use a validated sterilization procedure ●...
  • Page 36: Disinfection Components

    Cleaning , Disinfection and Sterilization 8.4 Disinfection Components Disinfection components Motor Handpiece Charge Base Adapter Handpiece Base Measuring Wire Wipe all the surfaces with a cloth lightly moistened with Ethanol for Disinfection (Ethanol 70 to 80vol%) at least 2 min, repeat for 5 times. NOTE Do not use anything except Ethanol for Disinfection (Ethanol 70 to 80 vol%).
  • Page 37: Error Warning

    Error Warning 9.Error Warning When setting the torque limit as R.L or during reverse processing, the Overload warning may Ov erload appear on the screen, it means a large load Restart Motor happened greater than the motor force. Press the Main switch to restart motor. Overheat temperature motor...
  • Page 38: Troubleshooting

    Troubleshooting 10.Troubleshooting When trouble is found, check the following points before contacting your distributor. If none of these are applicable or the trouble is not remedied even after action has been taken, the product may have failed. Contact your distributor. Ref.
  • Page 39 Troubleshooting contra-angle. Motor is protected by system Check the error warning. 9 or broken. The measure wire connecting Check the connection. not properly. Motor does not The lip hook not properly run when the file Check the connection. hooked in the corner of the is inserted in the patient’s mouth.
  • Page 40 Troubleshooting If it is the expected Beep sound an The motor is set to REV or mode, ignore the alarm. alarm even R.L mode. though the instrument is not being used. Canal Check situation of root Complex root canal measurement is canals.
  • Page 41: Technical Data

    Technical Data 11.Technical Data Manufacturer Changzhou Sifary Medical Technology Co.,Ltd Model E-connect S Dimensions 21.5cm x 17.5cm x 9cm±1cm (Outer box) Weight 1kg±10% Contra-angle compatible with rotary and reciprocating instruments, equipped with 2.35 shaft Contra-angle conforming to ISO 1797-1:2011, Type 1, Files length 11-31mm.
  • Page 42: Emc Tables

    Guidance and manufacturer’s declaration – electromagnetic emissions The E-connect S is intended for use in the electromagnetic environment specified below. The customer or the user of the E-connect S should assure that it is used in such an environment. Emissions test...
  • Page 43 Guidance and manufacturer’s declaration – electromagnetic immunity The E-connect S is intended for use in the electromagnetic environment specified below. The customer or the user of the E-connect S should assure that it is used in such an environment. IEC 60601 test...
  • Page 44 The E-connect S has been tested with the immunity test level in the below table and meet the related requirements of IEC 60601-1-2:2014. The customer and/or user should help keep a minimum distance between RF wireless communications equipments and the E-connect S as recommended below.
  • Page 45 Adapter Cable Measuring Wire 2. Use of E-connect S adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, E-connect S and the other equipment should be observed to verify that they are operating normally.
  • Page 46: Statement

    Statement 13.Statement Service Life The service life of E-connect S series products is 3 years. Maintenance MANUFACTURER will provide circuit diagrams, component part lists, descriptions, calibration instructions to assist to SERVICE PERSONNEL in parts repair. Disposal The package should be recycled. Metal parts of the device are disposed as scrap metal.
  • Page 47 Statement Changzhou Sifary Medcial Technology Co., Ltd Add: NO.99, Qingyang Road, Xuejia County, Xinbei District, Changzhou City, 213000 Jiangsu, China Tel: +86-0519-85962691 Fax: +86-0519-85962691 Email: ivy@sifary.com Web: www.sifary.com Llins Service & Consulting GmbH Tel: +49 175 4870819 Add: Obere Seegasse 34/2, 69124, Heidelberg, Germany Email:...