Fast-Pack
USER MANUAL

Advertisement

Table of Contents
loading

Summary of Contents for Eighteeth Fast-Pack

  • Page 1 Fast-Pack USER MANUAL...
  • Page 3: Table Of Contents

    Content 1. Scope of Fast-Pack ........................... 4 Parts Identification ......................4 Components and Accessories ..................5 Options (sold separately) ....................5 2. Symbols used in the User Manual ....................6 3. Before Use ............................7 Intended Use ........................7 Contraindications ......................7 4.
  • Page 4: Scope Of Fast-Pack

    Scope of Fast-Pack 1. Scope of Fast-Pack 1.1 Parts Identification 1.Charge Base 2.Handpiece 3.Heating Needle(3PCS) 4.Adapter Page 4 / 25...
  • Page 5: Components And Accessories

    Scope of Fast-Pack 1.2 Components and Accessories Fast-Pack Handpiece (1pcs) Charge Base (1pcs) Adapter (1pcs) Part No. 6351001 Part No. 6351003 Part No. 6316001+ 6316005 Heating needle S (1pcs) Heating needle M (1pcs) Heating needle L (1pcs) Size: 40/0.025 Size: 50/0.05 Size: 60/0.06...
  • Page 6: Symbols Used In The User Manual

    Symbols used in the User Manual 2. Symbols used in the User Manual If the instructions are not followed properly, operation may WARNING lead to hazards for the product or the user/patient. Additional information, explanation of operation and NOTE performance. Serial number Catalogue number Manufacturer...
  • Page 7: Before Use

    Before Use 3. Before Use 3.1 Intended Use Fast-Pack is intended for heating and cutting Gutta-percha outside the mouth during root canal treatment. This device must only be used in hospital environments, clinics or dental offices by qualified dental personnel.
  • Page 8: Installing The Fast-Pack

    Installing the Fast-Pack 4. Installing the Fast-Pack 4.1 Installation of heating WARNING needle ➢ To remove the heating needle, wait Make sure one of the dents on heating for it to cool down, it takes about 2-3 needle aligning the small triangle on seconds, and real-time temperature handpiece, push till to position.
  • Page 9: Connecting Charge Base

    Installing the Fast-Pack 4.3 Connecting charge base Plug the USB of adapter into the charge base, and plug the other end into a power outlet. NOTE Put the handpiece into the charge base in the right direction, otherwise the handpiece will not be charged.
  • Page 10: Use Interface

    Use Interface 5.Use Interface Turn Power On Long press Change Memory Shot press to change temperature memory from T1 to T5 Memory Parameter Setting During standby state, holding down press then press in 2 seconds to entry memory parameter setting, parameter of T1 to T5 can be set independently Heating Long press...
  • Page 11: Setting

    Setting 6.Setting 6.1 Memory Parameter Setting Fast-Pack has 5 memory programs, press 90℃ change during standby state, the memory number T1 will change according. During any Memory, holding down press then in 2 seconds, the “Temperature” of this press memory can be change.
  • Page 12: Advanced Setting

    Setting 6.2 Advanced Setting During power off state, holding down press then press to entry advanced setting, the “AutoPowerOff” will appear on the display screen. A ut o P ow e r Of f M in Press to adjust, the auto power off time can be set 5, 10 and 15 minutes press to confirm.
  • Page 13: Operation

    Operation 7.Operation 7.1 Charge Displays the present remaining amount of the battery. Less than 15% remains, please charge. NOTE If the power if less than 15%, must be recharged within 30 days, otherwise the battery will be damaged. Charge without charge base also available, using adapter connect to handpiece directly, the charge state will show on the screen.
  • Page 14: Heating

    Operation 7.2 Heating Press and hold the raised ridge on the ring switch to heat the heating needle. The indicator LED lights up during heating. Only the end of heating needle (about 4-5mm) will be heated, use this area to cut the gutta percha.
  • Page 15 Operation Release your finger from the ring switch, the heating needle will be cooling. the cooling indication will appear. 6 6 ℃ Real time temperature of cooling heating needle will appear. When the set time of “CoolingDisplay” is reached, will change to standby interface. heating needle isn’t installed...
  • Page 16: Maintenance

    Maintenance 8.Maintenance Autoclavable Components Heating needle Autoclavable Procedure Cleaning Disinfection Storage Packing Sterilization Cleaning: Clean the components with running water with a soft cloth. Disinfection: Wipe the components with a piece of gauze that has been dampened with Ethanol for Disinfection (Ethanol 70 to 80 vol%) and wrung out thoroughly.
  • Page 17 Maintenance Disinfection components Motor Handpiece Charge Base Adapter Handpiece Base Wipe the components with a piece of gauze that has been dampened with Ethanol for Disinfection (Ethanol 70 to 80 vol%) and wrung out thoroughly. NOTE Do not use anything except Ethanol for Disinfection (Ethanol 70 to 80 vol%) Do not use too much ethanol as it’s going into machine and damage the components inside.
  • Page 18: Troubleshooting

    Troubleshooting 9.Troubleshooting When trouble is found, check the following points before contacting your distributor. If none of these are applicable or the trouble is not remedied even after action has been taken, the product may have failed. Contact your distributor. Ref.
  • Page 19: Technical Data

    Technical Data 10.Technical Data Manufacturer Changzhou Sifary Medical Technology Co.,Ltd Model Fast-Pack Dimensions 21.5cm x 17.5cm x 9cm±1cm (Outer box) Weight 1kg±10% Lithium ion battery: 3.63V, 2600mAh, ±10% Power supply AC 100-240 V, ±10% Charger power supply Charger power output 50/60Hz, ±10%...
  • Page 20: Emc Tables

    11.EMC Tables Guidance and manufacturer’s declaration – electromagnetic emissions The Fast-Pack is intended for use in the electromagnetic environment specified below. The customer or the user of the Fast-Pack should assure that it is used in such an environment. Emissions test...
  • Page 21 Guidance and manufacturer’s declaration – electromagnetic immunity The Fast-Pack is intended for use in the electromagnetic environment specified below. The customer or the user of the Fast-Pack should assure that it is used in such an environment. Immuni...
  • Page 22 RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Fast-Pack is used exceeds the applicable RF compliance level above, the Fast-Pack should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting of relocating the Fast-Pack.
  • Page 23 Recommended separation distances between portable and mobile RF communications equipment and the Fast-Pack. The Fast-Pack is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Fast-Pack can help prevent...
  • Page 24: Statement

    Statement 12.Statement The service life of Fast-Pack series products is 3 years. All rights of modifying the product are reserved to the manufacturer without further notice. The pictures are only for reference. The final interpretation rights belong to CHANGZHOU SIFARY MEDICAL TECHNOLOGY CO., LTD. The industrial design, inner structure, etc, have claimed for several patents by SIFARY, any copy or fake product must take legal responsibilities.
  • Page 25 Changzhou Sifary Medcial Technology Co., Ltd Add: NO.99, Qingyang Road, Xuejia County, Xinbei District, Changzhou City, 213000 Jiangsu, China Tel: +86-0519-85962691 Fax: +86-0519-85962691 Email: ivy@sifary.com Web: www.eighteeth.com Landlink GmbH Tel: 0049 7641 9626855 Fax: 0049 7641 9626855 Add: Dorfstrasse 2/4, 79312, Emmendingen, Germany Email: info@landlink.eu Version: S01 IFU.010/S01...