Conducted dis-
turbances
induced by RF
fields
IEC 61000-4-6
Radiated RF
EM fields
IEC 61000-4-3
.
Proximity fields
from RF
wireless
communication
equipment
IEC 61000-4-3
Recommended minimum separation distances
Nowadays, many RF wireless equipments have being used in various healthcare
locations where medical equipment and/or systems are used. When they are used in
close proximity to medical equipment and/or systems, the medical equipment and/or
systems' basic safety and essential performance may be affected. The Fast-Pack pro
has been tested with the immunity test level in the below table and meet the related
3 V
0.15 MHz – 80
MHz, 6 V in ISM
bands and amateur
radio bands be-
tween 0.15 MHz
and 80 MHz, 80 %
AM at 1 kHz
3 V/m, 80 MHz –
2,7 GHz, 80 % AM
at 1 kHz
See the RF
wireless
communication
equipment table in
"Recommended
minimum
separation
distances"
3 V
3V/m
Complies
31 / 36
Portable and mobile RF
communications equipment
should be usedno closer to
any part of the Fast-Pack
pro, including cables, than
the recommended
separation distance
calculated from the
equation applicable to the
frequency of the
transmitter.
Recommended minimum
separation distances
See the RF wireless
communication equipment
table in "Recommended
minimum separation
distances"