Voltage
interruptions
IEC 61000-4-
11
Rated Power
frequency
magnetic field
IEC 61000-4-8
Note: UT: rated voltage(s); E.g. 25/30 cycles means 25 cycles at 50Hz or 30 cycles at 60Hz
Guidance and manufacturer's declaration – electromagnetic immunity
The Fast-Pack pro is intended for use in the electromagnetic environment specified below.
The customer or the user of the Fast-Pack pro should assure that it is used in such an
environment.
Immunity test
and 70% UT;
25/30 cycles
sine phase at 0°
0%
UT;
250/300 cycle
30 A/m
50Hz or 60Hz
IEC 60601 test
level
25/30 cycles
sine phase at 0°
0% UT; 250/300
cycle
30 A/m
50Hz or 60Hz
Compli
Electromagnetic environment -
ance
guidance
level
30 / 36
be powered form an
uninterruptible power
supply or a battery
Power frequency magnetic
field should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.