Eighteeth HyperLight User Manual

Eighteeth HyperLight User Manual

Hand-held dental x-ray system

Advertisement

Model: HyperLight
Hand-held Dental X-ray System
USER MANUAL
Changzhou Sifary Medical Technology Co.,Ltd.

Advertisement

Table of Contents
loading

Summary of Contents for Eighteeth HyperLight

  • Page 1 Model: HyperLight Hand-held Dental X-ray System USER MANUAL Changzhou Sifary Medical Technology Co.,Ltd.
  • Page 2 Version: S01 IFU-7035001 Issued: 2021.6.1 Size:197mmX140mm...
  • Page 3: Table Of Contents

    Content 1. Overview ........................4 1.1 Content............................4 1.2 Packing list and coding ........................5 2. Symbol instruction ......................6 3. Foreword ........................7 3.1 Scope of application ........................7 3.2 Contraindications ..........................7 4. Safety Precautions ......................9 4.1Radiation Safety ..........................9 4.2 Leakage Dose ..........................
  • Page 4: Overview

    1 Overview 1. Overview 1.1 Content 1.HyperLight Main Body 2.Base 3.Power Adapter with removable power cord 4.Backscatter Radiation Shielding Page 4 / 35...
  • Page 5: Packing List And Coding

    1 Overview 1.2 Packing list and coding HyperLight Main Body Power Adapter with Base(1 pcs) (1 pcs) removable power cord (1 pcs) Backscatter Radiation Shielding(option) Page 5 / 35...
  • Page 6: Symbol Instruction

    2 Symbol Instruction 2. Symbol Instruction If the instructions are not followed properly, operation WARNING may lead to hazards for the product or the user/patient. Additional information, explanation of operation and NOTE performance. Serial number Type BF applied part Alternating current Dispose of in accordance with the WEEE directive Keep dry Temperature limitation...
  • Page 7: Foreword

    3. Foreword 3.1 Scope of application The HyperLight Hand-held Dental X-ray System is indicated for use only by a trained and qualified dentist or dental technician for both adult and pediatric subjects as an extraoral diagnostic dental X-ray source to produce X-ray images using intraoral image receptors.
  • Page 8 3 Foreword Read the following warnings before use: 1. The device must not be placed in humid surroundings or anywhere where it can come into contact with any type of liquids. 2. Do not expose the device to direct or indirect heat sources. The device must be operated and stored in a safe environment.
  • Page 9: Safety Precautions

    4. Safety Precautions 4.1Radiation Safety The HyperLight was designed to be used in clinical settings (e.g., a dental office) and controlled settings where transportation or use of other X- ray devices might be prohibitive due to the device’s size and/or mobility.
  • Page 10: Leakage Dose

    4 Safety Precautions 4.2 Leakage Dose Leakage Dose Permissive Range 65 kVp, 2.5 mA, 0.5 s (Max. Exposure Condition) < 0.25 mGy/h At Focal Spot to Distance 1 m 1 : 30 Duty Cycle Page 10 / 35...
  • Page 11: Installation

    5 Installation 5. Installation 5.1 Plug in the charging cable First open the silicone cover, then plug one end of the charger into the machine and the other end into the socket. The charging icon will be displayed on the screen at this time. External extension cord interface Charging port Extension cords are not provided.
  • Page 12: Operation

    6 Operation 6 Operation 6.1 Operation panel instructions Symbol Type Choose adult mode or child mode Choose tooth type Increase exposure time Reduce exposure time Press two buttons at the same time to lock or unlock the machine. Press these two buttons at the same time to enter the setting interface.
  • Page 13: Poweron/Off

    6 Operation 6.2 Power On/Off 6.2.1 Power On 1. First press the power button and then let go, then the screen will enter the boot interface. 2. Make sure that the battery level is not lower than the figure below. Battery level 1 6.2.2 Power Off Long press the power button to shut down.
  • Page 14: Screen Instructions

    6 Operation 6.3 Screen instructions Item Description Bitewing Lower Molar Upper Molar Selects the tooth type. Canine Incisor Use the sensor to receive X- Digital Sensor rays Phosphor Use the Phosphor Plate to Plate receive X-rays Film Use the Film to receive X-rays Page 14 / 35...
  • Page 15 6 Operation The machine is locked and Machine lock cannot be exposed. Indicates remaining battery Remaining level. Battery When the indicator light starts Indicator to red glow, it means the battery needs to be charged. Battery Indicates the battery charger is Charging connected to the device.
  • Page 16: Set Interface

    6 Operation 6.4 Set interface After starting up, you can press the buttons at the same time to enter the setting interface, press the to turn the page, and press the and the to adjust the displayed parameters. Display content Meaning Adjust the volume of the machine.
  • Page 17: Use Of Exposure Function

    6 Operation 6.5 Use of exposure function 1. After selecting the exposure time, short press the exposure button to enter the ready state.  During the exposure, keep pressing the exposure button until the exposure is over. During the exposure, if you release the exposure button, the exposure will stop immediately.
  • Page 18: Positioning Instructions

    6 Operation 4. When the screen display is finish, it means that the exposure has been completed. Positioning Instructions 6.5.1 Paralleling technique The sensor is placed in a holder which is used to align the sensor parallel to the long axis of the teeth. Page 18 / 35...
  • Page 19: Bisected Angle Technique

    6 Operation 6.5.2 Bisected angle technique The patient holds the sensor in place with his/her finger. The X-ray beam is directed perpendicularly towards an imaginary line, which bisects the angle between the sensor plane and the long axis of the tooth. Position the tube head to the patient according to the accepted standard positioning procedures.
  • Page 20: Recommended Angle And Default Exposure Time For Each Tooth Type

    6 Operation 6.5.3 Recommended angle and default exposure time for each tooth type Angle of inclination(The Patie Exposure Receptor Teeth patient sits in a time(s) chair vertically.) +5˚~ +8˚ Bitewing Lower -5˚ Molar Upper Adult +30˚ 0.32 Molar Maxilla: +45˚ Canine 0.25 Mandible: -20˚...
  • Page 21 6 Operation Maxilla: +45˚ Canine 0.32 Mandible: -20˚ Maxilla: +45˚ Incisor Mandible: -25˚ +5˚~ +8˚ Bitewing Lower -5˚ 0.16 Molar Upper Child +30˚ 0.25 Molar Maxilla: +45˚ Canine 0.13 Mandible: -20˚ Maxilla: +45˚ Incisor Mandible: -25˚ +5˚~ +8˚ Bitewing Lower -5˚...
  • Page 22: Cleaning

    HyperLight plastic prematurely. 2. Unplug the charging cradle before cleaning. The HyperLight, the base, the charging cradle, and the AC power supply are not designed to be subjected to any kind of sterilization procedure. Page...
  • Page 23: Maintenance

    8 Maintenance 8 Maintenance If you do not use the machine for a long time, you need to charge the machine every 6 months. Page 23 / 35...
  • Page 24: Troubleshooting

    9 Troubleshooting 9 Troubleshooting When trouble is found, check the following points before contacting your distributor. If none of these are applicable or the trouble is not remedied even after action has been taken, the product may have failed. Contact your distributor.
  • Page 25: Technical Data

    10 Technical Data 10 Technical Data 10.1 Main Body Technical Data Changzhou Sifary Medical Technology Manufacturer Co., Ltd. Model HyperLight Package Dimensions 489mm×367.5mm×219mm(±10%) Total Weight 4kg±20% Charger power supply ~100-240V 50/60Hz Power supply Lithium ion battery: 14.8V, 2500mAh +10% Charger power output 1.5A...
  • Page 26 10 Technical Data Operating altitude < 3000m above sea level • Ambient temperature: -20 ° C ~ +55 ° C • Relative humidity: 20% - 80 %, non- Transport and storage condensing at > 40 ° C conditions • Atmospheric pressure: 70 kPa - 106 Page 26 / 35...
  • Page 27: X-Ray Tube Specifications And Characteristics

    10 Technical Data 10.2 X-ray Tube Specifications and Characteristics Filament voltage 2.4-3.0V Maximum filament current 2.9A Nominal tube voltage 70kV X-ray source assembly 4500J maximum heat content Maximum anode cooling rate 110W Nominal anode input power 600W Target material Tungsten Minimum target angle 12°...
  • Page 28 10 Technical Data : Page 28 / 35...
  • Page 29: Emc Tables

    11 EMC Tables 11 EMC Tables Guidance and manufacturer’s declaration – electromagnetic emissions HyperLight is intended for use in the electromagnetic environment specified below. HyperLight The customer or the user of the should assure that it is used in such an environment.
  • Page 30 Guidance and Manufacturer’s Declaration – Electromagnetic Emissions The HyperLight is intended for use in the electromagnetic environment specified below. The customer or user of the HyperLight should ensure that it is used in such an environment. Immunity IEC 60601...
  • Page 31 Guidance and Manufacturer’s Declaration – Electromagnetic Emissions The HyperLight is intended for use in the electromagnetic environment specified below. The customer or user of the HyperLight should ensure that it is used in such an environment. Electromagnetic Environment – Immunity...
  • Page 32 Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the HyperLight device The HyperLight is intended for use in the electromagnetic environment in which radiated disturbances are controlled. The customer or user of the HyperLight can help prevent...
  • Page 33 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz. Use of accessories and cables other than those specified or provided by the manufacturer HyperLight could result in increased electromagnetic emissions or decreased HyperLight electromagnetic immunity of and result in improper operation.
  • Page 34: Statement

    12 Statement 12 Statement Service Life The service life of HyperLight series products is 5 years. Maintenance MANUFACTURE will provide circuit diagrams, component part lists, descriptions, calibration instructions to assist to SERVICE PERSONNEL in parts repair. Disposal The package should be recycled. Metal parts of the device are disposed as scrap metal.
  • Page 35 12 Statement Changzhou Sifary Medical Technology Co., Ltd Add: NO.99, Qingyang Road, Xuejia County, Xinbei District, Changzhou City, 213000 Jiangsu, P. R. China Tel: +86-0519-85962691 Fax: +86-0519-85962691 Email: ivy@sifary.com Web: www.eighteeth.com Page 35 / 35...