Download Print this page

Endress+Hauser Dosimass IO-Link Operating Instructions Manual page 40

Coriolis flowmeter

Advertisement

Diagnostics and troubleshooting
40
Diagnostic
Short text
number
311
Electronic module faulty
331
Firmware update failed in
module 1 to n
372
Electronic module faulty
374
Electronic module faulty
Diagnostic of configuration
410
Data transfer failed
412
Processing download
419
Power cycle required
437
Configuration
incompatible
438
Dataset different
442
Frequency output 1
saturated
443
Pulse output 1 saturated
453
Flow override active
484
Failure mode simulation
active
485
Process variable
simulation active
492
Frequency output 1
simulation active
493
Pulse output 1 simulation
active
494
Switch output 1
simulation active
495
Diagnostic event
simulation active
Diagnostic of process
834
Process temperature too
high
835
Process temperature too
low
842
Process value below limit
862
Partly filled pipe
880
Output overloaded
910
Tubes not oscillating
Remedy instructions
Maintenance required!
Do not reset device
1. Update firmware of device
2. Restart device
1. Restart device
2. Check if failure recurs
3. Replace device
Restart device
1. Retry data transfer
2. Check connection
Download active, please wait
Power cycle device
1. Update firmware
2. Execute factory reset
1. Check dataset file
2. Check device parameterization
3. Download new device
parameterization
1. Check frequency output
settings
2. Check process
1. Check pulse output settings
2. Check process
Deactivate flow override
Deactivate simulation
Deactivate simulation
Deactivate simulation frequency
output
Deactivate simulation pulse output C
Deactivate simulation switch
output
Deactivate simulation
Reduce process temperature
Increase process temperature
Low flow cut off active!
Check low flow cut off
configuration
1. Check for gas in process
2. Adjust detection limits
Reduce load at the outputs
1. Check electronic module
2. Check sensor
Dosimass IO-Link
Status
Diagnostic
signal
behavior
[from the
[from the
factory]
factory]
M
Warning
F
Warning
F
Alarm
1)
S
Warning
F
Alarm
C
Warning
F
Alarm
F
Alarm
M
Warning
1)
S
Warning
1)
S
Warning
C
Warning
C
Alarm
C
Warning
C
Warning
Warning
C
Warning
C
Warning
1)
S
Warning
1)
S
Warning
1)
S
Warning
1)
S
Warning
S
Warning
F
Alarm