Error Messages and Remedies

Error messages
BANK DATA
Error in the bank data
ERROR
HEAD IS NOT
The Sensor Head is not connected cor-
CONNECTED
rectly.
NEIGHBOR UNIT
The Amplifier Units are not coupled cor-
IS NOT CON-
rectly.
NECTED
SYSDATA
Error in the system data.
ERROR
SYSTEM ERROR
Failed to configure FPGA.
ERROR CODE ??
Failed to initialize LCD.
Failed to recognize Amplifier Unit.
Failed to load data from flash memory.
Faulty hardware operation
Faulty software operation
TEACHING
The workpiece is not projected correctly.
FAILED
The teaching area is not set at the appro-
priate position.
In the following instances, error messages are not displayed, but the ERROR signal turns
ON.
Cause
TRIG was input while ENABLE was OFF.
Failed to teach from external device.
Cause
The bank data will be initialized, and
must be reset.
Make sure that the Sensor Head is
connected correctly.
Make sure that the Amplifier Units are
connected correctly.
The system data will be initialized, and
must be reset.
Faulty Amplifier Unit
Contact your OMRON representative.
Set the area so that the workpiece is
projected in the field-of-view.
Make sure that the appropriate teach-
ing area is set.
Countermeasure
Wait until ENABLE is turned ON and then input TRIG.
Set the area so that the workpiece is projected in the field-
of-view.
Make sure that the appropriate teaching area is set.
Make sure that TRIG timing is appropriate in the workpiece
move teaching mode.
Error Messages and Remedies
Countermeasure
Section 4