Initial Control Seltings; Procedure Steps; Check External Z-Axis Operation; Check Calibrator Amplitude And Frequency - Tektronix 2235A Instruction Manual

Portable oscilloscope
Table of Contents

Advertisement

Performance Check Procedure
-
2235A lnstructlon
EXTERNAL Z-AXIS AND CALIBRATOR
Equipment Required (See Table 4-1):
Leveled Sine-Wave Generator
BNC T-Connector
Two 50-52 BNC Coaxial Cables
1 O X Probe (provided with instrument)
50-52 BNC Termination
INITIAL CONTROL SETTINGS
PROCEDURE STEPS
Vertical
Channel 1 POSITION
MODE
BW LIMIT
CH 1 VOLTSIDIV
CH 1 VOLTSIDIV Variable
Input Coupling
Horizontal
POSITION
MODE
A SECIDIV
SECIDIV Variable
XI 0 Magnifier
A TRIGGER
VAR HOLDOFF
MODE
SLOPE
LEVEL
A 81 B SOLIRCE
A COUPL
Midrange
CH 1
On (button in)
1
v
CAL detent
DC
Midrange
A
10 J l
s
CAL detent
Off (knob in)
NORM
P-P AUTO
OUT
Midrange
VERT MODE
NORM
Check External Z-AXIS Operation
Connect the leveled sine-wave generator output via
a 50-52 cable and a T-connector to the CH
1
input
connector. Then connect a 50-52 cable and a 50-52
termination from the T-connector to the
EXT Z
AXIS
INPUT connector on the rear panel.
Set the generator to produce a 5-V, 50-kHz signal.
CHECK- For noticeable intensity modulation. The
positive part of the sine wave should be of lower
intensity than the negative part.
Disconnect the test equipment from the instrument.
Check Calibrator Amplitude and Frequency
Set:
CH 1 VOLTSIDIV
A SECIDIV
Connect the 10X Probe to the CH 1
OR
X input
connector and connect the probe tip to the CAL
connector.
Center the vertical display and, if necessary, adjust
the probe compensation for a flat top square-wave
display.
CHECK- For a 5-division display.
CHECK- For a 5-division period (5-divisions
between risetime portions of the display).
Disconnect the 1 O X probe from the instrument.