5. Verify the flange bolts are protruding through the sensor
module before applying pressure (see
Example
Table 2-1: Torque Values for the Flange and Flange Adapter Bolts
Bolt
material
Carbon Steel
(CS)
Stainless
Steel (SST)
Figure 2-5: Proper Bolt Installation
A. Bolt
B. Sensor module
10
Head markings
 
316
B8M
STM
316
316
R
 
Figure
Initial
torque
300 in-lb
B7M
150 in-lb
316
SW
316
A
B
January 2024
2-5).
Final torque
650 in-lb
300 in-lb
Emerson.com/Rosemount