CLEANING, DISINFECTION & MAINTENANCE
WARNING: RISK OF ELECTRIC SHOCK. ALWAYS DISCONNECT
THE POWER CORD FROM THE WALL AND THE INSTRUMENT
BEFORE PERFORMING ANY OF THE FOLLOWING CARE AND
MAINTENANCE PROCEDURES.

EXTERNAL CLEANING

Clean the external surfaces of this instrument using a clean,
soft cloth moistened with a mild detergent solution
of liquid dish soap (filtered below 5 microns). Refer to figure
20.

FOREHEAD/CHIN REST PREPARATION

For hygienic reasons, wipe the forehead rest with an alcohol
wipe and change the chin rest papers after each patient.
CLEANING THE GUIDE PLATE
If the Guide Plate is dirty it may cause a rough feeling when Maneuvering the base of the slit lamp. Clean
the Guide Plate with a sift cloth lightly dampened with a mild soap and water solution.
USER ENVIRONMENT REQUIREMENT
• Maintain Operational Environment as specified in Specification section of this manual.
• Do not use this Instrument in corrosive environment to prevent rusting etc.
CLEANING & DISINFECTION
Following disinfectants are recommended for cleaning and disinfection.
1. Normal household bleach (Sodium hypochlorite 5%)- strength 5000 ppm(10 parts water 1 part bleach).
2. 70% Isopropyl alcohol.
Procedure:
• Take a muslin cloth.
• Moist it to feel wet
• Disinfect/clean the surface gently.
Note – 1. Use mask and Gloves while performing cleaning and disinfection.
2. While cleaning, muslin cloth should not be dripped wet to prevent seepage and rusting to running/bare parts.
3. Alcohol is flammable,its use as a surface disinfectant should be in well-ventilated spaces only.
8126300-795
eVO 350
Figure-20 Cleaning Main Unit
Issue 1.2
Printed on June,2020
eVO 350