Download Print this page

Emerson Rosemount 1066 Quick Start Manual page 37

Single channel transmitter
Hide thumbs Also See for Rosemount 1066:

Advertisement

February 2024
5.5 
Canada
5.5.1  CSA intrinsically safe
Certificate
Standards
Markings
5.6 
Europe
5.6.1  ATEX/UKEX intrinsically safe
ATEX Certificate
UKEX Certificate
Standards
Markings
Quick Start Guide
Tamb = -20 °C to +65 °C
Non-Incendive:
Class I, Division 2, Groups ABCD T4
Dust Ignition Proof:
Class II, III, Division 1 Groups EFG
Install per DWG 1400669
IP66, Type 4X Enclosure
70176165
C22.2 No. 25-M1966, C22.2 No. 94-M91, C22.2
No.142-M1987, C22.2 No. 157-M1992, C22.2 No.
213-M1987, C22.2 No. 60529:05, UL 50, Eleventh
Edition, UL 508, Seventeenth Edition, UL 913,
Seventh Edition, UL 1203, Fourth Edition, ANSI/ISA
Standard: 12.12.01–2013
Intrinsically Safe:
Class I, II, III Division 1, Groups ABCDEFG T4
Tamb = -20 °C to +65 °C
Non-Incendive:
Class I, Division 2, Groups ABCD T4
Dust Ignition Proof:
Class II, III, Division 1 Groups EFG
Install per DWG 1400669
IP66, Type 4X Enclosure
Baseefa11ATEX0195X
BAS21UKEX0390X
EN IEC 60079-0:2018
EN 60079-11:2012
II 1 G
Ex ia IIC T4 Ga
(-20 °C ≤ Ta ≤ +65 °C)
Quick Start Guide