Quick Start Guide
5. The transmitter case has six ½-in. (12.7 mm) conduit openings.
Use separate conduit openings for the power/output cable,
the sensor cable, and the other sensor cable as needed (pH
input for free chlorine with continuous pH correction).
6. Use weathertight cable glands to keep moisture out of the
transmitter. If using a conduit, plug and seal the connections
at the transmitter housing to prevent moisture from getting
inside the instrument.
7. Install cable gland fittings and plugs as needed to properly
seal the transmitter on all six enclosure openings. The USB
port cover must be fully installed on the front cover to ensure
proper transmitter sealing.
1.4 
Preparing conduit openings
There are six conduit openings in all configurations of the
transmitter.
Note
Emerson fits four of the openings with plugs upon shipment.
14
February 2024
Emerson.com/Rosemount