What to do if...
38
10.2.2

Error message OFL, UFL

The measured value is outside the measuring range of the meter.
The measured value is obviously incorrect.
Cause
• Cable broken
• Gel electrolyte dried out
• Air bubble in front of the junction
• Air in the junction
10.3

General information

10.3.1

Symbol for calibration evaluation flashes

Cause
Calibration interval expired
10.3.2

[LoBat] display

Cause
Batteries almost empty
10.3.3

Instrument does not react to keystroke

Cause
Operating condition undefined or
EMC load unallowed
Remedy
• Replace the electrode
• Replace the electrode
• Remove the air bubble
• Extract air or moisten the junction
Remedy
Recalibrate the measuring system
Remedy
Replace the batteries (see section
3.3.1 I
NSERTING THE BATTERIES
10)
Remedy
• Processor reset:
Press the <ENTER> and
<On/Off> key simultaneously
ba77143e02
Lab 855
, page
12/2017