What to do if...
38
10.2.2
Error message OFL, UFL
The measured value is outside the measuring range of the meter.
The measured value is obviously incorrect.
Cause
• Cable broken
• Gel electrolyte dried out
• Air bubble in front of the junction
• Air in the junction
10.3
General information
10.3.1
Symbol for calibration evaluation flashes
Cause
Calibration interval expired
10.3.2
[LoBat] display
Cause
Batteries almost empty
10.3.3
Instrument does not react to keystroke
Cause
Operating condition undefined or
EMC load unallowed
Remedy
• Replace the electrode
• Replace the electrode
• Remove the air bubble
• Extract air or moisten the junction
Remedy
Recalibrate the measuring system
Remedy
Replace the batteries (see section
3.3.1 I
NSERTING THE BATTERIES
10)
Remedy
• Processor reset:
Press the <ENTER> and
<On/Off> key simultaneously
ba77143e02