pH value
22
3.
If necessary, press <CAL> to select the buffer set ([ConCal]).
ASY is displayed.
4.
Thoroughly rinse the pH electrode with deionized water.
5.
Immerse the pH electrode in the first buffer solution (pH 7.0 ± 0.5 for
two-point calibration).
6.
When measuring without temperature sensor: Enter the temperature of
the buffer with <><>.
7.
Start the measurement with <ENTER> .
The pH value of the buffer solution is displayed.
The measured value is checked for stability (stability control).
The [AR] status display flashes.
8.
Wait for the measurement with stability control to be completed.
9.
Set the nominal pH value of the buffer solution with <><>.
10.
Accept the calibration value with <ENTER>.
SLO is displayed.
11.
If necessary, finish the calibration procedure as a single-point calibration
with <M>.
The calibration record is displayed.
or
Continue calibration using the next buffer with <ENTER>.
Continuing with two-point calibration
12.
Thoroughly rinse the pH electrode with deionized water.
13.
Immerse the pH electrode in the second buffer solution.
A
S
Y
2
4
8
°C
ConCal
s
L
O
2
8
4
°C
ConCal
Lab 855
TP
TP
ba77143e02
12/2017