Technical data
11
Technical data
40
11.1
Measuring ranges, resolution, accuracy
11.1.1
Measuring ranges, resolution
Parameter
pH
U [mV]
T [°C]
T [°F]
11.1.2
Manual temperature input
Parameter
T
[°C]
manual
T
[°F]
manual
11.1.3
Accuracy (± 1 digit)
Parameter
*
pH / range
- 2.0 ... + 20.0
- 2.00 ... + 20.00
- 2.000 ... + 19.999
U [mV] / range
- 2000 ... + 2000
-1200.0 ... +1200.0
T [°C] / temperature sensor
• NTC 30
• PT 1000
* when measuring in a range of ± 2 pH around a calibration point
Measuring range
- 2.0 ... + 20.0
- 2.00 ... + 20.00
- 2.000 ... + 19.999
- 1200.0 ... + 1200.0
- 2000 ... + 2000
- 5.0 ... + 105.0
+ 23.0 ... + 221.0
Range
- 25 ... + 130
-13 ... + 266
Accuracy
± 0.1
± 0.01
± 0.005
± 1
± 0.3
± 0.1
± 0.1
Resolution
0.1
0.01
0.001
0.1
1
0.1
0.1
Increment
1
1
Temperature of the test sample
+ 15 °C ... + 35 °C
+ 15 °C ... + 35 °C
+ 15 °C ... + 35 °C
+ 15 °C ... + 35 °C
+ 15 °C ... + 35 °C
ba77143e02
Lab 855
12/2017
Need help?
Do you have a question about the SI Analytics Lab 855 and is the answer not in the manual?
Questions and answers