Buffer sets for calibration
You can use the buffer sets quoted in the table for an automatic calibration. The
pH values are valid for the specified temperature values. The temperature
dependence of the pH values is taken into account during the calibration.
The buffer set is selected during calibration (see section 5.2.3
A
UTOMATIC CALIBRATION
No.
1
TEC
Technical buffers
2
NIST/DIN
DIN buffers according to DIN 19266
and NIST Traceable Buffers
3
ConCal
Possible setting
(Change with
<><>)
Resolution of pH display
Temperature unit
Calibration interval
(A
C
UTO
Buffer set
Description
), page 19).
AL
pH values
2,000
25 °C
4,010
7,000
10,011
1.679
25 °C
4.006
6.865
9.180
12.454
1-point cal.:
any,
• Any
adjust-
able
2-point cal.:
• 7.0 ± 0.5
• Any
Settings
at