● Do not use bleach or chloride disinfectant materials.
6.3 Cleaning disinfection sterilization
Autoclavable Components
Sleeve
Only the components above can be autoclaved.
Before first use and after each use, clean disinfect and sterilize the above
components.
Reprocessing Instructions
Before cleaning, disconnect the components from the Main unit. Refer
to Chapter 4.3 of this manual for disassembly instructions. Remove
gross contaminations from the components with cold water (<40° C)
immediately after use. Don't use a fixating detergent or hot water
(>40° C) as this can cause the fixation of residuals which may influence
the result of the reprocessing process.
Preparat
Store the components in a humid surrounding.
ion
at
the Point
of Use:
Do not submerge the components or wipe them with any of the
following functional water (acidic electrolyzed water, strong alkaline
solution, or ozone water), medical agents (glutaral, etc.), or any other
special types of water or commercial cleaning liquids. Such liquids may
result in metal corrosion and adhesion of the residual medical agents
to the components.
Transpor
Safe storage and transportation to the reprocessing area to avoid any
tation:
damage and contamination to the environment.
Preparat
The devices must be reprocessed in a disassembled state.
ion
for
Deconta
Observe suitable personal protective measures.
mination
:
Pre-
Do a manual pre-cleaning, until the components are visually clean.
Handpiece
holder
30