be operated and stored in a safe environment.
 The device requires special precautions with regard to electromagnetic
compatibility (EMC) and must be installed and operated in strict compliance with
the EMC information. In particular, do not use the device in the vicinity of
fluorescent lamps, radio transmitters, remote controls and do not use this system
near the active HF Surgical Equipment in the hospital. Portable RF
communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm (12 inches) to any part
of the Dental Anesthetic Delivery System, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment could
result.
 Do not charge, operate or store at high temperatures. Comply with the
specified operating and storage conditions.
 Gloves are compulsory during the operation.
 If irregularities occur in the device during treatment, switch it off. Contact the
agency.
 Never open or repair the device yourself, otherwise, void the warranty.
 Do not use the device while charging.
 Do not place the device in a position where it might fall in case of injury to
patients or damage to the device.
 When battery leakage occurs, handle the leakage according to local laws and
regulations to avoid environmental pollution.
 The device should not be used in environments where flammable materials
are present.
 Do not spray alcohol on the connection interface.
 Do not use conductive objects to detect the connection interface.