2) In "PDL mode", when the displayed "pressure
value" is bigger than the value set in the system,
the prompt tone "beep" becomes higher, and the
green cell of the "pressure bar chart" is light up.
3) When the system pressure is too high, the
system will sound a sharp prompt tone, and the last
orange cell of "pressure bar chart" will light up, and
the system will automatically stop injection. After
the pressure is released, press the foot pedal again
to continue the injection operation.
5.3.8 Smart reminder
When "smart reminder" is enabled (click the icon
with your finger to turn this function on
and off), the system will add the following prompt
tone:
1) When the dosage of anesthetic is less than 20%,
the system will prompt "remain low"; When the
dosage is exhausted, the system will prompt
"cartridge empty ".
2) In PDL mode, the first green cell of "pressure bar
chart" lights up, and the system will prompt "PDL".
3) When the system pressure is too high, the
25
It is recommended to
enable
reminder ".